Product

Allura Xper systems with Certeray X-ray Generator located in regions with high mains (480V) voltage. Product Usage: The Allura Xper series is intended for use on human patients to perform: Vascular, cardiovascular and neurovascular imaging applications, including diagnostic, interventional and minimally invasive procedures. This includes, e.g., peripheral, cerebral, thoracic and abdominal angiography, as well as PTAs, stent placements, embolisations and thrombolysis. Cardiac imaging applications including diagnostics, interventional and minimally invasive procedures (such as PTCA, stent placing, atherectomies), pacemaker implanatations, and electrophysiology (EP). Non-vascular interventions such as drainages, biopsies and vertebroplasties procedures.

FDA product code
OWBInterventional Fluoroscopic X-Ray System
Device class
Class 2
Medical specialty
Radiology
510(k) numbers
K133292, K141979
Affected lot / code info
Allura Xper systems - Model #'s 722026, 722027, 722028, 722029, 722033, 722034, 722035, 722036, 722038, 722039, 722058, 722059, 722063, 722064, 722065, 722066, 722078, 722079, (added as of 6/25/2018) 1120

Why it was recalled

There is an increase in the failure rate of certain Anode Drive Units (ADU5) used in these products. Early life failure of the ADUs only occurs when the hospital mains provide 480V to the system. This current voltage may lead to saturation and overheating of the coils that protect the IGBTs (insulated-gate bipolar transistor) of the ADU. The overheating may also generate a peculiar burning smell that may be noticed. In some instances this burning smell may be noticed before the ADU fails. When the ADU fails the X-ray performance of the system is reduced to Emergency Fluoroscopy. Exposure will not be possible and image quality is reduced.

Root cause (FDA determination)

Device Design

Action the firm took

Customer Notification letter dated August 2017 were sent to affected customers. The letter identified affected product, problem and actions to be taken. The customer is requested to place this Field Safety Notice with the documentation of the system until this correction has been implemented. For questions contact your local Philips representative.

Recalling firm

Firm
Philips Electronics North America Corporation
Address
3000 Minuteman Rd, Andover, Massachusetts 01810-1032

Distribution

Distribution pattern
Worldwide Distribution - US Nationwide; foreign distribution to Brazil, Germany, Iceland, Japan, Mexico, and (updated 6/25/18) Saudi Arabia.

Timeline

Recall initiated
2018-01-04
Posted by FDA
2018-02-24
Terminated
2019-10-09
Status

Source: openFDA Device Recall endpoint. Recall record ID #161339. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.