Recalls / —
—#162165
Product
MINI PLASTICS SET
- FDA product code
- LRP — Tray, Surgical
- Device class
- Class 1
- Medical specialty
- General, Plastic Surgery
- Affected lot / code info
- 75700 Lot codes: 2015102790 2016011590 2016030790 2016050490 2016062390 2016071590 2016081790 2016111890 2017012090 2017020790 2017042090
Why it was recalled
Convenience Kit contains a PVP Prep Pad, received by a supplier, that does not meet the iodine assay level required to support the 36 month shelf life.
Root cause (FDA determination)
Nonconforming Material/Component
Action the firm took
On October 16, 2017 a PRODUCT CORRECTION letter was issued to customers notifying them that if they had any kits in their inventory, Centurion will supply labels to affix to the front of each kit instructing the user to discard the prep pad upon opening the kit. Other components of the kit may continue to be used and are NOT within the scope of this recall.
Recalling firm
- Firm
- Centurion Medical Products Corporation
- Address
- 100 Centurion Way, Williamston, Michigan 48895-9086
Distribution
- Distribution pattern
- Nationally
Timeline
- Recall initiated
- 2017-10-23
- Terminated
- 2018-06-21
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #162165. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.