Recalls / —
—#162503
Product
Dimension Vista Thyroid Stimulating Hormone (TSH) Flex reagent cartridge
- FDA product code
- JLW — Radioimmunoassay, Thyroid-Stimulating Hormone
- Device class
- Class 2
- Medical specialty
- Clinical Chemistry
- 510(k) numbers
- K140859
- Affected lot / code info
- TSH, radioimmunoassay, thyroid stimulating hormone, RF612, SMN# 10464524 Lot codes: 16277BD 16322BB 17005BB 17023AC 17069AA 17093BB 17122AA 17150AA 17193BA 17215AC 17268AA
Why it was recalled
The concentrations for biotin listed in the non-interfering substances section of the instructions for use for the Dimension TSHL, Dimension LOCI FT4L, Dimension Vista TSH, ADVIA Centaur Cyclosporine, and ADVIA Centaur DHEAS, incorrectly state the level at which biotin does not interfere.
Root cause (FDA determination)
Error in labeling
Action the firm took
Siemens Healthcare Diagnostics issued a recall in March 2017 for products with incorrect units for biotin in the Instructions for Use. After further investigation, recalling firm determined that more Siemens products also had missing or incorrect information in the IFUs for which two recalls were initiated, and Amended Corrections and Removals Reports were submitted to the FDA
Recalling firm
- Firm
- Siemens Healthcare Diagnostics, Inc.
- Address
- 511 Benedict Ave, Tarrytown, New York 10591-5005
Distribution
- Distribution pattern
- Nationwide and Canada, Mexico, and Thailand
Timeline
- Recall initiated
- 2018-01-02
- Posted by FDA
- 2018-05-04
- Terminated
- 2023-02-10
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #162503. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.