Recalls / —
—#162958
Product
ADVIA Centaur CA 125II Assay; Cat No. 09427226, SMN 10310443 (1-pack/100 tests); Cat no. 01678114, SMN 10315696 (5-pack/500 tests)
- FDA product code
- LTK — Test, Epithelial Ovarian Tumor-Associated Antigen (Ca125)
- Device class
- Class 2
- Medical specialty
- Immunology
- 510(k) numbers
- K020828
- Affected lot / code info
- SMN 10310443 (1-pack/100 tests): Lot 71804174, Exp 2018-Apr-27, UDI (01)00630414461298(10)71804174(17)20180427 Lot 73838174, Exp 2018-Apr-27, UDI (01)00630414461298(10)73838174(17)20180427 Lot 88601174, Exp 2018-Apr-27, UDI (01)00630414461298(10)88601174(17)20180427 Lot 94549174, Exp 2018-Apr-27, UDI (01)00630414461298(10)94549174(17)20180427 Lot 23904176, Exp 2018-Aug-30, UDI (01)00630414461298(10)23904176(17)20180830 Lot 24439176, Exp 2018-Aug-30, UDI (01)00630414461298(10)24439176(17)20180830 Lot 31842176, Exp 2018-Aug-30, UDI (01)00630414461298(10)31842176(17)20180830 Lot 42051176, Exp 2018-Aug-30, UDI (01)00630414461298(10)42051176(17)20180830 Lot 54008176, Exp 2018-Aug-30, UDI (01)00630414461298(10)54008176(17)20180830 SMN 10315696 (5-pack/500 tests): Lot 71229174, Exp 2018-Apr-27, UDI (01)00630414461304(10)71229174(17)20180427 Lot 73455174, Exp 2018-Apr-27, UDI (01)00630414461304(10)73455174(17)20180427 Lot 84113174, Exp 2018-Apr-27, UDI (01)00630414461304(10)84113174(17)20180427 Lot 91568174, Exp 2018-Apr-27, UDI (01)00630414461304(10)91568174(17)20180427 Lot 23903176, Exp 2018-Aug-30, UDI (01)00630414461304(10)23903176(17)20180830 Lot 24944176, Exp 2018-Aug-30, UDI (01)00630414461304(10)24944176(17)20180830 Lot 41442176, Exp 2018-Aug-30, UDI (01)00630414461304(10)41442176(17)20180830 Lot 46735176, Exp 2018-Aug-30, UDI (01)00630414461304(10)46735176(17)20180830
Why it was recalled
Upon dilution, some patient samples may exhibit over-recovery outside the representative data provided in the assay instructions for use.
Root cause (FDA determination)
Other
Action the firm took
A Customer Notification was sent to customers on 2/21/18. The letter instructs customers to refer to the information provided until the appropriate updates are available in the assay instructions for use. The letter is to be retained and forwarded to those who may have received the product.
Recalling firm
- Firm
- Siemens Healthcare Diagnostics, Inc
- Address
- 333 Coney St, East Walpole, Massachusetts 02032-1516
Distribution
- Distribution pattern
- US Nationwide, and the following countries: Albania, Angola, Argentina, Armenia, Australia, Austria, Bahrain, Bangladesh, Belgium, Bosnia Herzegovina, Brazil, Bulgaria, Cambodia, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Ecuador, Egypt, Estonia, Finland, France, Georgia, Germany, Greece, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Latvia, Lithuania, Luxembourg, Macedonia, Malaysia, Malta, Mexico, Morocco, Myamar, Nepal, Netherlands, New Zealand, Nicaragua, Norway, Pakistan, Paraguay, Peru, Philippines, Poland, Portugal, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sweden, Switzerland, Taiwan,Thailand, Turkey, United Arab Emirates, Ukraine, United Kingdom, Uruguay, Uzbekistan,Vatican City, and Vietnam.
Timeline
- Recall initiated
- 2018-02-21
- Terminated
- 2019-05-16
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #162958. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.