FDA Device Recalls

Recalls /

#163454

Product

Connecting Tube - Female to Male Luer Lock, HPCT8.8-60-M-FM-BNS Product Usage: The connecting tube is intended to be a conduit between a fluid reservoir and a vascular catheter in the patient. The connecting tube may be used during angiographic procedures to connect radiographic contrast power injector to a vascular catheter, arterial pressure monitoring, and fluid infusion.

FDA product code
DTLAdaptor, Stopcock, Manifold, Fitting, Cardiopulmonary Bypass
Device class
Class 2
Medical specialty
Cardiovascular
Affected lot / code info
NS7881928, NS7898133

Why it was recalled

Cap-adapter assembly of the device may potentially not be adequately tightened, possibly leading to fluid leakage from or air aspiration into the connecting tube.

Root cause (FDA determination)

Equipment maintenance

Action the firm took

On March 29, 2018, the recalling firm mailed recall notifications to affected customers. Customers were informed that the affected devices may have been manufactured with equipment that was out of calibration, which could result in the cap-adapter assembly of the device not being adequately tightened. Customers were advised of potential adverse events. Customers were instructed to 1) Examine inventory immediately to determine if they have affected product, and quarantine affected product. Immediately cease all distribution and use of the recalled lots. 2) Return the affected product to the recalling firm with a copy of the Acknowledgement and Receipt Form to receive a product credit. 3) Complete the Acknowledgement and Receipt Form even if the customer does not have affected products on hand. 4) Immediately report adverse events to the recalling firm. Customers with further questions were directed to contact Cook Medical Customer Relations at 800-457-4500 or 812-339-2235.

Recalling firm

Firm
Cook Inc.
Address
750 N Daniels Way, Bloomington, Indiana 47404-9120

Distribution

Distribution pattern
Distributed to accounts in 7 states: AL, CA, GA, NE, NV, NY, and TX. International distribution to Japan, Korea, and Taiwan.

Timeline

Recall initiated
2018-03-29
Posted by FDA
2018-05-09
Terminated
2019-04-09
Status

Source: openFDA Device Recall endpoint. Recall record ID #163454. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.