Recalls / —
—#163719
Product
POWERPORT ClearVUE Slim with Smooth Septum, 6F ChronoFlex Silk, Polyurethane Catheter, REF 5676300
- FDA product code
- LJT — Port & Catheter, Implanted, Subcutaneous, Intravascular
- Device class
- Class 2
- Medical specialty
- General Hospital
- 510(k) numbers
- K122899
- Affected lot / code info
- Lot Number Unique Device Identifier (UDI) REAN0055 (01)00801741110771(17)170628(10)REAN0055 REAN0202 (01)00801741110771(17)170528(10)REAN0202 REAP0771 (01)00801741110771(17)170828(10)REAP0771 REAS1708 (01)00801741110771(17)171028(10)REAS1708 REAT0147 (01)00801741110771(17)171128(10)REAT0147 REAT1852 (01)00801741110771(17)171228(10)REAT1852 REAU1811 (01)00801741110771(17)171228(10)REAU1811 REAU2420 (01)00801741110771(17)171228(10)REAU2420 REAV2465 (01)00801741110771(17)180228(10)REAV2465 REAW1422 (01)00801741110771(17)180331(10)REAW1422 REAW1831 (01)00801741110771(17)180331(10)REAW1831 REAX1132 (01)00801741110771(17)180331(10)REAX1132 REAY1623 (01)00801741110771(17)180331(10)REAY1623 REAY2597 (01)00801741110771(17)180331(10)REAY2597 REBP0090 (01)00801741110771(17)180331(10)REBP0090 REBS1269 (01)00801741110771(17)181031(10)REBS1269 REBR1002 (01)00801741110771(17)181031(10)REBR1002 REBS1439 (01)00801741110771(17)181031(10)REBS1439 REBT1052 (01)00801741110771(17)181231(10)REBT1052 REBT2348 (01)00801741110771(17)181231(10)REBT2348 REBV0714 (01)00801741110771(17)190131(10)REBV0714 REBV1758 (01)00801741110771(17)190131(10)REBV1758 REBW1288 (01)00801741110771(17)190331(10)REBW1288 REBX1628 (01)00801741110771(17)190331(10)REBX1628 REBX2052 (01)00801741110771(17)190430(10)REBX2052
Why it was recalled
The Patient ID Card packaged with the POWERPORT ClearVUE Implantable Port with Smooth Septum contains incorrect port flushing instructions.
Root cause (FDA determination)
Error in labeling
Action the firm took
The firm initiated the recall by letter on 02/08/2018. The letter explained the problem and the products involved and requested the following actions: "Please check all inventory locations within your institution for affected product code / lot number combinations listed in the field correction notice. If you have further distributed any of the product code / lot numbers, please immediately contact that location, advise them of this field correction notice, and forward these instructions. For any identified product on your shelves, please replace the products Patient ID Card with the corrected Patient ID Card enclosed with this field correction notice. Enclosed with this field correction notice is a Patient Communication Packet that contains one patient letter, one corrected Patient ID Card, and one POWERPORT Patient Guide. Please contact all patients that have received any product listed in this field correction notice and provide them with the Patient Communication Packet."
Recalling firm
- Firm
- Bard Peripheral Vascular Inc
- Address
- 1625 W 3rd St, Bldg 1, Tempe, Arizona 85281-2438
Distribution
- Distribution pattern
- Nationwide
Timeline
- Recall initiated
- 2018-02-08
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #163719. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.