Product

POWERPORT ClearVUE Slim with Smooth Septum, 6F ChronoFlex Silk, Polyurethane Catheter with Open Suture Holes, REF 5676301

FDA product code
LJTPort & Catheter, Implanted, Subcutaneous, Intravascular
Device class
Class 2
Medical specialty
General Hospital
510(k) numbers
K122899
Affected lot / code info
Lot Number Unique Device Identifier (UDI) REAN0066 (01)00801741110795(17)170528(10)REAN0066 REAP0629 (01)00801741110795(17)170728(10)REAP0629 REAY1483 (01)00801741110795(17)180531(10)REAY1483 REAY2494 (01)00801741110795(17)180331(10)REAY2494 REBS0163 (01)00801741110795(17)181031(10)REBS0163 REBT0364 (01)00801741110795(17)181130(10)REBT0364 REBT1664 (01)00801741110795(17)181231(10)REBT1664 REBT2149 (01)00801741110795(17)181231(10)REBT2149 REBW0874 (01)00801741110795(17)190331(10)REBW0874 REBY1231 (01)00801741110795(17)190430(10)REBY1231

Why it was recalled

The Patient ID Card packaged with the POWERPORT ClearVUE Implantable Port with Smooth Septum contains incorrect port flushing instructions.

Root cause (FDA determination)

Error in labeling

Action the firm took

The firm initiated the recall by letter on 02/08/2018. The letter explained the problem and the products involved and requested the following actions: "Please check all inventory locations within your institution for affected product code / lot number combinations listed in the field correction notice. If you have further distributed any of the product code / lot numbers, please immediately contact that location, advise them of this field correction notice, and forward these instructions. For any identified product on your shelves, please replace the products Patient ID Card with the corrected Patient ID Card enclosed with this field correction notice. Enclosed with this field correction notice is a Patient Communication Packet that contains one patient letter, one corrected Patient ID Card, and one POWERPORT Patient Guide. Please contact all patients that have received any product listed in this field correction notice and provide them with the Patient Communication Packet."

Recalling firm

Firm
Bard Peripheral Vascular Inc
Address
1625 W 3rd St, Bldg 1, Tempe, Arizona 85281-2438

Distribution

Distribution pattern
Nationwide

Timeline

Recall initiated
2018-02-08
Status