Product
POWERPORT ClearVUE Slim Implantable Port with Smooth Septum and Attachable 8F Polyurethane Open-Ended Single-Lumen Venous Catheter Custom Kit, REF CP00005
- FDA product code
- LJT — Port & Catheter, Implanted, Subcutaneous, Intravascular
- Device class
- Class 2
- Medical specialty
- General Hospital
- 510(k) numbers
- K122899
- Affected lot / code info
- Lot Number Unique Device Identifier (UDI) REAN0056 (01)00801741110894(17)170628(10)REAN0056 REAN0203 (01)00801741110894(17)170628(10)REAN0203 REAP0773 (01)00801741110894(17)170728(10)REAP0773 REBN0925 (01)00801741110894(17)180131(10)REBN0925
Why it was recalled
The Patient ID Card packaged with the POWERPORT ClearVUE Implantable Port with Smooth Septum contains incorrect port flushing instructions.
Root cause (FDA determination)
Error in labeling
Action the firm took
The firm initiated the recall by letter on 02/08/2018. The letter explained the problem and the products involved and requested the following actions: "Please check all inventory locations within your institution for affected product code / lot number combinations listed in the field correction notice. If you have further distributed any of the product code / lot numbers, please immediately contact that location, advise them of this field correction notice, and forward these instructions. For any identified product on your shelves, please replace the products Patient ID Card with the corrected Patient ID Card enclosed with this field correction notice. Enclosed with this field correction notice is a Patient Communication Packet that contains one patient letter, one corrected Patient ID Card, and one POWERPORT Patient Guide. Please contact all patients that have received any product listed in this field correction notice and provide them with the Patient Communication Packet."
Recalling firm
- Firm
- Bard Peripheral Vascular Inc
- Address
- 1625 W 3rd St, Bldg 1, Tempe, Arizona 85281-2438
Distribution
- Distribution pattern
- Nationwide
Timeline
- Recall initiated
- 2018-02-08
- Status
- —