Recalls / —
—#163785
Product
smith&nephew 5o (5 degree) GENESIS(R) II VALGUS BUSHING, REF CATALOG NUMBER 71440014 The GENESIS II Knee System Components are indicated for rheumatoid arthritis; post-traumatic arthritis, osteoarthritis, or degenerative arthritis in older patients whose age, weight, and activity levels are compatible with an adequate long-term result; failed osteotomies, unicompartmental replacement, or total knee replacement. The constrained knee systems are designed for use in patients in primary and revision surgery, where the posterior cruciate ligament and one or both of the collateral ligaments (i.e., medial collateral and/or lateral collateral ligament) are absent or incompetent.
- FDA product code
- MBH — Prosthesis, Knee, Patello/Femorotibial, Semi-Constrained, Uncemented, Porous, Coated, Polymer/Metal/Polymer
- Device class
- Class 2
- Medical specialty
- Orthopedic
- 510(k) numbers
- K121393
- Affected lot / code info
- Lot number 16MM00441
Why it was recalled
A single lot of GENESIS II FEMORAL FIVE DEGREE VALGUS BUSHING due to a laser mark labeling error. The left valgus bushing was incorrectly laser marked as right and vice versa.
Root cause (FDA determination)
Process control
Action the firm took
The firm sent an Urgent Medical Device Recall Notice dated April 9, 2018. Required Actions: 1. Please inspect your inventory and locate any unused devices from the listed product and batch numbers on the first page of this Field Action Notification, and quarantine them immediately. a. If you are a distributor, you must notify your customers of the field action and ensure that these actions are carried out. 2. If you have no product to return, please put an X in the appropriate location below. 3. If you have product to return, please list the batch numbers and quantities of each batch that you are returning in the appropriate boxes below. 4. Complete the remainders of the form sign and send to FieldActions@smith-nephew.com or fax to 901-566- 7975. Please Note even if you have no product to return, this form must be completed, signed and returned. 5. Once the form is received by Smith & Nephew, you will be sent a Return Authorization (RA) number. If you have any questions or concerns regarding this recall please contact FieldActions@smith-nephew.com. For further questions, please call (978) 749-1440.
Recalling firm
- Firm
- Smith & Nephew, Inc.
- Address
- 1450 E Brooks Rd, Memphis, Tennessee 38116-1804
Distribution
- Distribution pattern
- US Distribution to the states of : NC, IN, CO MI, PA, NY, MA, CA, GA
Timeline
- Recall initiated
- 2018-04-09
- Terminated
- 2019-02-22
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #163785. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.