FDA Device Recalls

Recalls /

#163899

Product

Pressure Injectable ARROWg+ard Blue PLUS(R) Multi-Lumen Central Venous Catheterization Kit; Product Codes: ASK-42703-PAM1, ASK-42703-PSHR1, ASK-45703-PNY, ASK-45703-PWHC1

FDA product code
FOZCatheter, Intravascular, Therapeutic, Short-Term Less Than 30 Days
Device class
Class 2
Medical specialty
General Hospital
510(k) numbers
K071538
Affected lot / code info
Lot Numbers:  Product Code ASK-42703-PAM1: 13F17K0077 13F17L0494   Product code ASK-42703-PSHR1: 13F17C0361 13F17F0207 13F17H0518 13F17M0278   Product code ASK-45703-PNY: 13F17H0124 13F17H0451   Product code ASK-45703-PWHC1: 13F17E0022

Why it was recalled

Product sterility may be compromised due to unsealed packaging.

Root cause (FDA determination)

Under Investigation by firm

Action the firm took

On April 11, 2018 Arrow International mailed Urgent Medical Device Recall notifications to affected customers. Customers were instructed to take the following actions: 1. If you have affected stock, immediately discontinue use and quarantine any products with the product codes and lot numbers listed above, so that the affected products can be returned to Arrow. 2. To return product, complete the enclosed Recall Acknowledgement Form and fax it to 1-855-419-8507, Attn: Customer Service or email to recalls@teleflex.com. This will allow us to document the amount of product you have on hand for return. A customer service representative will contact you with a Return Goods Authorization (RGA) Number and will provide instructions for the return of products to Arrow International. 3. If you have no affected stock, please complete the enclosed Recall Acknowledgment Form and fax it to 1-855-419-8507, Attn: Customer Service or email to recalls@teleflex.com. This will allow us to document your receipt of this letter. Customers with questions may contact your local sales representative or Customer Service at (866)246-6990.

Recalling firm

Firm
Arrow International Inc
Address
2400 Bernville Rd, Reading, Pennsylvania 19605-9607

Distribution

Distribution pattern
US Nationwide, AUSTRALIA, AUSTRIA, BELGIUM, CANADA, CZECH REPUBLIC, DENMARK, ECUADOR, FRANCE, GERMANY, GREECE, HONG KONG, INDONESIA, IRELAND, ITALY, JAPAN, KOREA, NETHERLANDS, NEW ZEALAND, PANAMA, PHILIPPINES, POLAND, PUERTO RICO, ROMANIA, SAUDI ARABIA, SINGAPORE, SLOVENIA, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, THAILAND, TRINIDAD AND TOBAGO, UNITED KINGDOM

Timeline

Recall initiated
2018-04-11
Terminated
2020-05-15
Status

Source: openFDA Device Recall endpoint. Recall record ID #163899. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.