Product

IntelliVue MX40 Patient Monitor, Software Versions B.05, B.06, B.06.5X, Model 865352

FDA product code
DSIDetector And Alarm, Arrhythmia
Device class
Class 2
Medical specialty
Cardiovascular
510(k) numbers
K113125
Affected lot / code info
All devices since release of B.05 (March 2016) prefix date code >/= US096

Why it was recalled

Five warning statements are missing from the instructions for use.

Root cause (FDA determination)

Error in labeling

Action the firm took

Urgent Medical Device Correction notification letters were distributed on 4/30/18. The notification included the following instructions: Customers were asked to please review the information provided in the letter with all members of their staff who need to be aware of its contents. The letter provides information on the missing warnings and instructions for actions to be taken. The enclosed MX40 IFU Errata Sheet must be attached to the first page of Chapter 6 of the Instructions for Use for ready reference. Complete and return the attached Customer Reply Form.

Recalling firm

Firm
Philips Electronics North America Corporation
Address
3000 Minuteman Rd, Andover, Massachusetts 01810-1032

Distribution

Distribution pattern
Worldwide Distribution -- US, to the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, and to the territories of Guam and Puerto Rico; and, countries of Argentina, Australia, Austria, Belgium, Canada, Chile, China, Croatia, Czech Republic, Denmark, Estonia, Finland, France, German, Ghana, Greece, Guadeloupe, Hong Kong, Iceland, India, Ireland, Israel, Italy, Japan, Latvia, Lebanon, Malaysia, Martinique, Mexico, Myanmar, Netherlands, New Zealand, Norway, Philippines, Poland, Portugal, Qatar, Russia, Saudi Arabia, Singapore, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, UK, UAE, and Vietnam.

Timeline

Recall initiated
2018-04-30
Terminated
2019-10-04
Status