Product
Biomet Modular Primary Tibial Tray Implants; Modular Interlok Primary Tray (size 75 mm); Item No. 141215; UDI: (01) 00880304005310 (17) 280306 (10) 385340 Product is used in Total Knee Arthroplasty
- FDA product code
- JWH — Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer
- Device class
- Class 2
- Medical specialty
- Orthopedic
- 510(k) numbers
- K010212, K142933, K171054, K915132
- Affected lot / code info
- Lot 385340 UDI: (01) 00880304005310 (17) 280306 (10) 385340
Why it was recalled
One lot of 75mm tibial tray is potentially etched and labeled as a 79 mm tibial tray. This may lead to delay in surgery, or immediate revision surgery to remove an incorrectly sized tibial component after it has been cemented.
Root cause (FDA determination)
Under Investigation by firm
Action the firm took
On May 18, 2018, the firm notified affected distributors and user facilities of the recall through Urgent Medical Device Recall letters sent via email and FedEx. The letters informed affected consignees of the issue and stated that the affected units were distributed in March 2018. Distributor Responsibilities: 1. Review this notification and ensure that affected team members are aware of the contents. 2. Immediately locate and quarantine affected product in your inventory. 3. Immediately return all affected product from your distributorship and from affected hospitals within your territory. a. Complete Attachment 1 Inventory Return Certification Form and send to CorporateQuality.PostMarket@zimmerbiomet.com within three (3) days. b. For each return, send a copy of Attachment 1 to CorporateQuality.PostMarket@zimmerbiomet.com. c. Include a hardcopy of Attachment 1 in each carton of your return shipment for immediate processing. d. Mark RECALL on the outside of the returned cartons. 4. Return the Additional Accounts form to CorporateQuality.PostMarket@zimmerbiomet.com. a. Review the list of hospitals and/or surgeons included with the email notification sent to your facility, which includes a list of hospitals and/or surgeons that have already been notified of this recall. b. Identify whether there are any additional hospitals and/or surgeons that Zimmer Biomet has not notified and list these accounts on the Additional Accounts form. Please provide the form in Excel format. c. If there are no additional accounts or surgeons to notify, please indicate that there are no additional accounts, or indicate None or NA on the form. Risk Manager Responsibilities: 1. Review this notification and ensure that affected personnel are aware of the contents. 2. If you have affected product at your facility, assist your Zimmer Biomet sales representative and quarantine all affected product. Your Zimmer Biomet sales representative will remove the affected
Recalling firm
- Firm
- Zimmer Biomet, Inc.
- Address
- 56 E Bell Dr, Warsaw, Indiana 46582-6989
Distribution
- Distribution pattern
- US Distribution to the states of : FL, GA, IN, MA, NM, UT, and WI. Foreign distribution to Canada and Mexico
Timeline
- Recall initiated
- 2018-05-18
- Posted by FDA
- 2018-06-01
- Terminated
- 2020-03-31
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #164798. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.