FDA Device Recalls

Recalls /

#165194

Product

da Vinci Xi EndoWrist Stapler 45 Reload Green, Model 48445G-03, 12 reloads/box.

FDA product code
GDWStaple, Implantable
Device class
Class 2
Medical specialty
General, Plastic Surgery
Affected lot / code info
UDI: 00886874112823. Device expiration: 10/14/2017 02/21/2020

Why it was recalled

Intuitive Surgical, Inc. (ISI) has recently become aware of an issue with the da Vinci Xi Stapler 45 Blue and Green Reloads that may be missing a component of the knife mechanism. Without this component, the reload knife will not move and cut tissue when the stapler is fired.

Root cause (FDA determination)

Nonconforming Material/Component

Action the firm took

Intuitive Surgical, Inc. (ISI) conducted the recall and notified customers via letter "Field Safety Notice Urgent Medical Device Notification" sent the week of April 23, 2018. The letter notified customers of the recall, reason for field action, risk to health, affected products, Actions to be taken by the Customer/User, Actions to be taken by Intuitive Surgical, and Further Information and Support: If you need further information or support concerning this Medical Device Notification , please contact your Clinical Sales Representative or contact Intuitive Surgical Customer Service at the numbers listed below: "North and South America: (800) 876-1310, Option 3 (4 AM to 5 PM PST) or mail: customerservice@intusurg.com. "Europe, Middle East, Asia and Africa: +800 0821 2020 or +41 21 821 2020 (8AM to 6 PM CET) or ics@intusurg.com "South Korea: 02-3271-3200 (9 AM to 6 PM KSTJ) "Japan: 0120-56-5635 or 03-5575-1362 (9 AM to 6 PM JST).

Recalling firm

Firm
Intuitive Surgical, Inc.
Address
1266 Kifer Rd, Bldg 101, Sunnyvale, California 94086-5304

Distribution

Distribution pattern
U.S., foreign and government accounts. Foreign: Australia, Austria, Belgium, Brazil, Canada, China (Hong Kong), Czech Republic, Denmark, Finland, France, Germany, Ireland, Italy, Japan, Luxembourg, Netherlands, Norway, Qatar, Saudi Arabia, Singapore, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Kingdom.

Timeline

Recall initiated
2018-04-23
Terminated
2020-07-15
Status