FDA Device Recalls

Recalls /

#165939

Product

Pediatric Two-Lumen Central Venous Catheterization Kit with Blue FlexTip ARROWg+ard Blue Catheter, Cat. No. AK-22502 Product Usage: The multiple-lumen catheter permits venous access to the central circulation of pediatric patients. The ARROWg+ard antimicrobial surface is intended to help provide protection against catheterrelated infections. The catheter is not intended to be used as a treatment for existing infections, nor is it indicated for long-term use. The use of the 4, 5, and 5.5 Fr. antimicrobial surface catheters in infants weighing less than 2.0, 2.5, and 3.0 kilograms respectively, have not been evaluated. Therefore, the benefits of these catheters in infants weighing less than the indicated weights should be weighed against any possible risks.

FDA product code
FOZCatheter, Intravascular, Therapeutic, Short-Term Less Than 30 Days
Device class
Class 2
Medical specialty
General Hospital
510(k) numbers
K900263
Affected lot / code info
13F17K0202

Why it was recalled

One lot of product is from product code AK-22502 so it contains a 2-lumen, 5 Fr x 5 cm catheter. However, this product is incorrectly labeled as AK-25502, so the lidstock says it contains a 2-lumen, 5 Fr x 8 cm catheter. No patient injuries have been reported related to this issue.

Root cause (FDA determination)

Employee error

Action the firm took

On July 3, 2018, the firm sent out Urgent Medical Device Recall letters to customers. The letter communicated the labeling error. Affected consignees are instructed to take the following actions: 1. If you have affected stock, immediately discontinue use and quarantine any products with the product code and lot number listed above, so that the affected products can be returned to Arrow. 2. To return product, complete the enclosed Recall Acknowledgement Form and fax it to 1-855-419-8507, Attn: Customer Service or email to recalls@teleflex.com. This will allow us to document the amount of product you have on hand for return. A customer service representative will contact you with a Return Goods Authorization (RGA) Number and will provide instructions for the return of products to Arrow International. 3. If you have no affected stock, please complete the enclosed Recall Acknowledgment Form and fax it to 1-855-419-8507, Attn: Customer Service or email to recalls@teleflex.com. This will allow us to document your receipt of this letter.

Recalling firm

Firm
Arrow International Inc
Address
2400 Bernville Rd, Reading, Pennsylvania 19605-9607

Distribution

Distribution pattern
Distributors in MO and UT. Medical facilities in AL, CA, IL, MA, NH, NM, NY, TX, VA, and WI.

Timeline

Recall initiated
2018-07-03
Terminated
2020-04-29
Status

Source: openFDA Device Recall endpoint. Recall record ID #165939. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.

— recall: Pediatric Two-Lumen Central Venous Catheterization Kit with Blue FlexTip ARROWg+ard Blue Catheter, C · FDA Device Recalls