Recalls / —
—#166904
Product
Vanguard 360 Revision Knee System 75x 5mm Universal Tibial Block Augment, Model Number 185224 Intended for use in Total Knee Arthroplasty
- FDA product code
- JWH — Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer
- Device class
- Class 2
- Medical specialty
- Orthopedic
- 510(k) numbers
- K093293
- Affected lot / code info
- Lot 585070, UDI (01) 0088030448157 (17) 230215 (10) 585070
Why it was recalled
The implant only has one bolt in the sterile package but it should have had two.
Root cause (FDA determination)
Under Investigation by firm
Action the firm took
The firm, Zimmer Biomet, sent an "URGENT MEDICAL DEVICE RECALL" notification letters dated 8/2/18 to customers on 8/2/18. The letters described the product, problem and actions to be taken. The letters instruct customers to perform the following actions: Risk Manager Responsibilities: 1. Review this notification and ensure that affected personnel are aware of the contents. 2. If you have affected product at your facility, assist your Zimmer Biomet sales representative and quarantine all affected product. Your Zimmer Biomet sales representative will remove the affected product from your facility. 3. Complete Attachment 1 Certificate of Acknowledgement and send to CorporateQuality.PostMarket@zimmerbiomet.com. This form must be returned even if you do not have affected products at your facility. 4. Retain a copy of the acknowledgement form with your recall records in the event of a compliance audit of your facilitys documentation. 5. If you have further questions or concerns after reviewing this notice, please call customer service at 574-371-3071 between 8:00 am and 5:00pm EST, Monday through Friday. Calls received outside of call center operating hours will receive a voicemail prompt or be transferred to an on-call representative in the event of an emergency. Alternatively, your questions may be emailed to CorporateQuality.PostMarket@zimmerbiomet.com. Distributors: 1. Review this notification and ensure that affected team members are aware of the contents. 2. Immediately locate and quarantine affected product in your inventory. 3. Immediately return all affected product from your distributorship and from affected hospitals within your territory. a. Complete Attachment 1 Inventory Return Certification Form and send to CorporateQuality.PostMarket@zimmerbiomet.com within three (3) days. b. For each return, send a copy of Attachment 1 to CorporateQuality.PostMarket@zimmerbiomet.com. c. Include a hardcopy of Attachment 1 in each carton of your retur
Recalling firm
- Firm
- Zimmer Biomet, Inc.
- Address
- 56 E Bell Dr, Warsaw, Indiana 46582-6989
Distribution
- Distribution pattern
- Worldwide Distribution: US (nationwide) to the following US states: AZ, KS, NM, NV, and UT; and to countries of: Australia and Netherlands.
Timeline
- Recall initiated
- 2018-08-02
- Terminated
- 2019-04-25
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #166904. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.