Recalls / —
—#167334
Product
1) Cytomics FC 500 With CXP Software BECKMAN COULTER 2) Cytomics FC 500 MPL With MXP Software BECKMAN COULTER flow cytometer for in vitro diagnostics
- FDA product code
- GKZ — Counter, Differential Cell
- Device class
- Class 2
- Medical specialty
- Hematology
- 510(k) numbers
- K030828, K071681
- Affected lot / code info
- FC 500 Catalog Num FC 500" Flow Cytometer MCL with UPS 626553 FC 500" Flow Cytometer MPL 626554 FC 500" MCL Flow Cytometer 5 Color (100V) 6605627 FC 500" MCL Flow Cytometer 5 Color (120V) 6605628 FC 500" MCL Flow Cytometer 5 Color (220V) 6605629 FC 500" MCL Flow Cytometer 5 Color (240V) 6605630 FC 500 Flow Cytometer, Remfg Assy, 5 Clr (220V) A89264 FC 500" Flow Cytometer 100-240V, 50-60HZ, Remfg Assy A89151 FC 500 Recond MCL UPS Flow Cytometer 100-240V, 50-60HZ A88262 FC 500 Flow Cytometer 100-240V A88263 All products manufactured affected for the time period: FC-500 From 01/22/2002 to 01/31/2018 EPICS XL/XL-MCL From 11/01/1992 to 03/15/2013
Why it was recalled
A manufacturing defect on the amplifier board of the Flow Cytometers my cause signal loss and/or signal drifting resulting in absence of data or a population shift in the data plots. This can impact patient results for any application.
Root cause (FDA determination)
Nonconforming Material/Component
Action the firm took
On 01/12/2018 the firm sent an Urgent Medical Device Recall Letter to affected customers via email/mail informing them that as a result of customer complaints and subsequent internal investigations, Beckman Coulter has determined that an internal electronic component on the circuit Amplifier boards in the FC 500 system may be affected by a manufacturing defect. Each FC 500 system contains seven (7) of the potentially affected Amplifier boards. All instruments are potentially impacted. Customers were informed to implement the following actions for the application used: 1. For All Applications including Laboratory Developed Tests: a. As per the product documentation all data must be reviewed by a laboratory professional prior to the release of reported results from the lab. b. Immediately implement the collection of Time as a parameter and create Time versus Parameter plots which will allow the monitoring of signal integrity during data acquisition as instructed in Attachment 2. c. Review data as described below: i. Review of all Time plots for each parameter. ii. Monitor consistent Forward Scatter, Side Scatter and all fluorescence data over time as shown in Attachment 2. iii. Unexpected fluctuations in the events over time may indicate compromised fluidics, signal integrity or data acquisition conditions. iv. All data must be reviewed prior to the release of any results from the laboratory, via an LIS or any other mechanism. 2. For tetraCXP and stemCXP: a. As it is not possible to add time versus parameter, discontinue use of the automated tetraCXP and stemCXP applications. b. You can continue to use the tetraCHROME reagents and Stem Kit reagents with the manual gating instructions provided in the product labeling. Refer to CYTO-STAT tetraCHROME IFU, PN B90108 (for tetraCHROME CD45- FITC/CD4-RD1/CD8-ECD/CD3-PC5, PN 6607013 and tetraCHROME CD45- FITC/CD56-RD1/CD19-ECD/CD3-PC5, PN 6607073) and the Stem-Kit IFU, PN B6
Recalling firm
- Firm
- Beckman Coulter Inc.
- Address
- 250 S Kraemer Blvd, Brea, California 92821-6232
Distribution
- Distribution pattern
- Worldwide distribution. US Nationwide including Puerto Rico. OUS: FC 500: Japan, South Korea, China, Italy, Spain, United Kingdom, Germany, Bangladesh, Czech Republic, France, Pakistan, India, South Africa, Austria, Saudi Arabia, Algeria, Turkey, Belgium, Netherlands, Bulgaria, Sweden, Australia, Hungary, Sri Lanka, Uruguay, Canada, New Zealand, Switzerland, Gabon, Portugal, Mexico, Singapore, Monaco, Hong Kong, Guadeloupe, Zambia, United Arab Emirates, Ireland, Reunion, Poland, Egypt, Slovakia, Croatia, Brazil, Chile, Bahrain, Slovenia, Bosnia and Herzegovina, Greece, Myanmar, Morocco, Serbia, Philippines, Lebanon, Romania, Moldova, Qatar, Barbados, Iraq, Oman, Namibia, Indonesia, Thailand, The Bahamas, Macau, El Salvador, Panama, Denmark, Venezuela, Azerbaijan, Israel, Colombia, Peru, Malaysia, Vietnam, Tunisia, Kuwait, Kenya, Libya, Latvia, Russia, Belarus, Ukraine, Kazakhstan, Taiwan EPICS XL: Albania, Algeria, Angola, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Botswana, Bulgaria, Burundi, Cameroon, Canada, China, Congo, Croatia, Cyprus, Czech Republic, Denmark, Dominican, Republic, Egypt, Ethiopia, France, Germany, Greece, Guinea, Guinea-Bissau, Hong Kong, India, Indonesia, Ireland, Israel, Italy, Japan, Kazakhstan, Kenya, Kuwait, Lebanon, Lesotho, Libya, Macau, Malawi, Mexico, Morocco, Mozambique, Namibia, Netherlands, New Zealand, Nigeria, Oman, Pakistan, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Singapore, Slovakia, South Africa, South Korea, Spain, Sri Lanka, Sudan, Swaziland, Sweden, Switzerland, Syria, Taiwan, Tanzania, The Bahamas, Tunisia, Turkey, Turkmenistan, Uganda, Ukraine, United Arab Emirates, United Kingdom, Zambia, Zimbabwe
Timeline
- Recall initiated
- 2018-01-12
- Posted by FDA
- 2018-11-20
- Terminated
- 2024-04-30
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #167334. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.