Product

VariCam

FDA product code
KPSSystem, Tomography, Computed, Emission
Device class
Class 2
Medical specialty
Radiology
510(k) numbers
K953801
Affected lot / code info
Serial Numbers: 0000000010-203 0000000010-191 000003100-1829 00008125-18-18 00000000010-15 00000000010-63 00008125-16-59 000000099-4-04 00000000010-31 00000000010-82 00008125-16-70 00008125-18-09 00008125-18-16 00000000017-64 10280 00000000010-30 000003681-1803 00000000018-24 000003100-1840 000003101-1502 000003100-1744 00000000017-36 00000000016-86 00000000017-31 1065 1998 1999 000008125-1630 1008 00000000017-20 00000000017-22 00000000017-23 1804 00000000030-14 00000000030-15 00000000010-37 00000000010-44 00000000010-55 00000000030-16 00000000018-39 1704 00000ICPO50612 1697 1696 000000163-1758 00000000018-23 00000000016-73 00000000030-04 00000000010-59 00000000018-22 000003100-1760 000003100-1743 0000000010-134 000003100-1719 343 000003100-1763 61061 00000000010-04 00000000030-11 1649 00000000010-67 00000000010-45 00000000017-27 00000000017-15

Why it was recalled

A potential hazardous situation may occur if the collimator locking handle is not properly placed in the lock position and the warning circuit does not detect that the collimator is unlocked because of the patient proximity to the collimator during a clinical scan.

Root cause (FDA determination)

Other

Action the firm took

The firm initiated the medical device correction by letter on 10/0/2017. The letter instructed the consignee to ensure the collimator locking handles are in the locked position at all times. GE Healthcare will inspect and if necessary correct all affected products.

Recalling firm

Firm
GE Healthcare, LLC
Address
3000 N Grandview Blvd, Waukesha, Wisconsin 53188-1615

Distribution

Distribution pattern
Worldwide

Timeline

Recall initiated
2017-10-30
Status

Source: openFDA Device Recall endpoint. Recall record ID #169654. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.