Recalls / —
—#170401
Product
Phoenix AP AST Indicator Bag, Catalog Number 246006
- FDA product code
- LON — System, Test, Automated, Antimicrobial Susceptibility, Short Incubation
- Device class
- Class 2
- Medical specialty
- Microbiology
- 510(k) numbers
- K050745
- Affected lot / code info
- Batch Numbers 8234684 8234533 8225579 8225578 8218607 8218605 8213696 8213635 8213633 8208581 8208579 8206585 8190531 8190530 8170557 8234505 8170562 8159691 8150670 8150669
Why it was recalled
Panels inoculated using certain lots of the indicator solution are demonstrating an increased occurrence of test aborts.
Root cause (FDA determination)
Under Investigation by firm
Action the firm took
Urgent Medical Device Product Advisory letters dated 12/20/18 were sent to customers. Amended Urgent Medical Device Product Advisory letters dated 1/23/19 were also sent to customers.
Recalling firm
- Firm
- Becton Dickinson & Co.
- Address
- 7 Loveton Cir, Sparks, Maryland 21152-9212
Distribution
- Distribution pattern
- The products were distributed to the following US states: AL, AR, AZ, CA, CO, CT, DC, FL, GA, IA, ID, IL, IN, KS, KY, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI, WV, and WY. The products were distributed to the following foreign countries: Argentina, Australia, Bangladesh, Belgium, Brazil, Canada, Cayman Islands, Chile, China, Columbia, Costa Rica, Dominican Republic, Ecuador, Guatemala, India, Indonesia, Japan, Korea, Malaysia, Mexico, Myanmar, New Zealand, Pakistan, Panama, Peru, Philippines, San Salvador, Singapore, Sri Lanka, Thailand, Taiwan, Uruguay, Vietnam
Timeline
- Recall initiated
- 2018-12-20
- Terminated
- 2020-04-10
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #170401. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.