#170625

Product

Angled Tip Ureteral Catheter Set with Bentson PTFE Wire Guide, Ureteral Catheter, Global Product No. G19108 G14587 G15302 G14326

FDA product code
OCYEndoscopic Guidewire, Gastroenterology-Urology
Device class
Class 2
Medical specialty
Gastroenterology, Urology
Affected lot / code info
8018728 8080823 8314382 8409523 8023194 7857489

Why it was recalled

The wire guide may be incorrectly loaded into the wire guide holder, leading to the stiff tip of the wire guide being introduced into the patient instead of the flexible tip.

Root cause (FDA determination)

Process control

Action the firm took

On January 4, 2019, the firm distributed Urgent Medical Device Recall letters to customers via courier mail. Customers were informed that affected lots may contain a wire guide that was incorrectly loaded into the wire guide holder.

Recalling firm

Firm
Cook Inc.
Address
750 N Daniels Way, Bloomington, Indiana 47404-9120

Distribution

Distribution pattern
Domestic distribution to nationwide US. Foreign distribution to Australia, Belgium, Bolivia, Canada, Canary Islands, Chile, China, Colombia, Croatia, Czech Republic, Denmark, France, French Polynesia, Germany, Guatemala, Honduras, Hungary, India, Ireland, Italy, Japan, Korea, Luxembourg, Malaysia, Malta, Mexico, Mongolia, Netherlands, New Caledonia, New Zealand, Oman, Panama, Paraguay, Poland, Puerto Rico, Romania, Russian Federation, South Africa, Spain, Sweden, Switzerland, Taiwan, Trinidad and Tobago, United Arab Emirates, and the United Kingdom.

Timeline

Recall initiated
2019-01-04
Terminated
2020-05-29
Status

Source: openFDA Device Recall endpoint. Recall record ID #170625. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.