Recalls / —
—#170636
Product
Sof-Flex Multi-Lenqth Ureteral Stent Set, Global Product No. G14773 G15000 G14867 G14951 G17852
- FDA product code
- FAD — Stent, Ureteral
- Device class
- Class 2
- Medical specialty
- Gastroenterology, Urology
- Affected lot / code info
- 8011038 NS7998389 NS8008388 NS8018651 NS8030247 NS8037667 7990842 8599605 8625597 8625598 8172131 8172132 8193001
Why it was recalled
The wire guide may be incorrectly loaded into the wire guide holder, leading to the stiff tip of the wire guide being introduced into the patient instead of the flexible tip.
Root cause (FDA determination)
Process control
Action the firm took
On January 4, 2019, the firm distributed Urgent Medical Device Recall letters to customers via courier mail. Customers were informed that affected lots may contain a wire guide that was incorrectly loaded into the wire guide holder.
Recalling firm
- Firm
- Cook Inc.
- Address
- 750 N Daniels Way, Bloomington, Indiana 47404-9120
Distribution
- Distribution pattern
- Domestic distribution to nationwide US. Foreign distribution to Australia, Belgium, Bolivia, Canada, Canary Islands, Chile, China, Colombia, Croatia, Czech Republic, Denmark, France, French Polynesia, Germany, Guatemala, Honduras, Hungary, India, Ireland, Italy, Japan, Korea, Luxembourg, Malaysia, Malta, Mexico, Mongolia, Netherlands, New Caledonia, New Zealand, Oman, Panama, Paraguay, Poland, Puerto Rico, Romania, Russian Federation, South Africa, Spain, Sweden, Switzerland, Taiwan, Trinidad and Tobago, United Arab Emirates, and the United Kingdom.
Timeline
- Recall initiated
- 2019-01-04
- Terminated
- 2020-05-29
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #170636. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.