FDA Device Recalls

Recalls /

#170636

Product

Sof-Flex Multi-Lenqth Ureteral Stent Set, Global Product No. G14773 G15000 G14867 G14951 G17852

FDA product code
FADStent, Ureteral
Device class
Class 2
Medical specialty
Gastroenterology, Urology
Affected lot / code info
8011038 NS7998389 NS8008388 NS8018651 NS8030247 NS8037667 7990842 8599605 8625597 8625598 8172131 8172132 8193001

Why it was recalled

The wire guide may be incorrectly loaded into the wire guide holder, leading to the stiff tip of the wire guide being introduced into the patient instead of the flexible tip.

Root cause (FDA determination)

Process control

Action the firm took

On January 4, 2019, the firm distributed Urgent Medical Device Recall letters to customers via courier mail. Customers were informed that affected lots may contain a wire guide that was incorrectly loaded into the wire guide holder.

Recalling firm

Firm
Cook Inc.
Address
750 N Daniels Way, Bloomington, Indiana 47404-9120

Distribution

Distribution pattern
Domestic distribution to nationwide US. Foreign distribution to Australia, Belgium, Bolivia, Canada, Canary Islands, Chile, China, Colombia, Croatia, Czech Republic, Denmark, France, French Polynesia, Germany, Guatemala, Honduras, Hungary, India, Ireland, Italy, Japan, Korea, Luxembourg, Malaysia, Malta, Mexico, Mongolia, Netherlands, New Caledonia, New Zealand, Oman, Panama, Paraguay, Poland, Puerto Rico, Romania, Russian Federation, South Africa, Spain, Sweden, Switzerland, Taiwan, Trinidad and Tobago, United Arab Emirates, and the United Kingdom.

Timeline

Recall initiated
2019-01-04
Terminated
2020-05-29
Status

Source: openFDA Device Recall endpoint. Recall record ID #170636. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.