FDA Device Recalls

Recalls /

#171914

Product

Discovery NM/CT 670 ES Product Usage: The GE Discovery NM/CT 670 system is a medical tool intended for use by appropriately trained health professionals to aid in detecting, localizing, diagnosing of diseases and organ function for the evaluation of diseases, trauma, abnormalities, and disorders such as, but not limited to, cardiovascular disease, neurological disorders and cancer. The system output can also be used by the physician for staging and restaging of tumors, planning, guiding, and monitoring therapy. The GE Discovery NM/CT 670 system, combining Nuclear Medicine (NM) and Computed Tomography (CT) systems, is intended to produce: The GE Discovery NM/CT 670 system is a medical tool intended for use by appropriately trained health professionals to aid in detecting, localizing, diagnosing of diseases and organ function for the evaluation of diseases, trauma, abnormalities, and disorders such as, but not limited to, cardiovascular disease, neurological disorders and cancer. The system output can also be used by the physician for staging and restaging of tumors, planning, guiding, and monitoring therapy. The GE Discovery NM/CT 670 system, combining Nuclear Medicine (NM) and Computed Tomography (CT) systems, is intended to produce:

FDA product code
JAKSystem, X-Ray, Tomography, Computed
Device class
Class 2
Medical specialty
Radiology
510(k) numbers
K093514
Affected lot / code info
Mfg. Lot or Serial # System ID Model, Catalogue or Code # ESDX35020 403777D670C H3100BL ESPX54001 A51773134 H3100BM ESPY54002 DE203284 H3100BM

Why it was recalled

During installation of certain Nuclear Medicine systems delivered with dismounted detectors, some detector screws may not have been secured correctly, leading to the possible loss of one detector mounting design redundancy.

Root cause (FDA determination)

Device Design

Action the firm took

GE Healthcare sent an Urgent Medical Device Correction letter dated March 20, 2019 to affected customers. The letter identified the affected product, problem and actions to be taken. GE Healthcare will inspect and, if required, correct all affected products at no cost to you to ensure that all design redundancies are in place. A GE Healthcare representative will contact you to arrange for the correction. For questions contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative.

Recalling firm

Firm
GE Healthcare, LLC
Address
3000 N Grandview Blvd, Waukesha, Wisconsin 53188-1615

Distribution

Distribution pattern
US and Austria, Canada, Croatia, France, Germany, India, Italy, Japan, Republic of Korea, Netherlands, Poland, Portugal, Slovakia, South Africa, Spain, Sweden, Switzerland, UK

Timeline

Recall initiated
2019-03-20
Terminated
2020-10-05
Status

Source: openFDA Device Recall endpoint. Recall record ID #171914. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.