Recalls / —
—#172802
Product
Maquet Getinge: (1) Material: 701053941R01 BEQ-TOP 9005 3/8 Custom Pk w/RF32 & HMOD (2) Material: 701053941R02- BEQ-TOP 9005 3/8 Custom Pk w RF32 & HMOD Material:
- FDA product code
- DWF — Catheter, Cannula And Tubing, Vascular, Cardiopulmonary Bypass
- Device class
- Class 2
- Medical specialty
- Cardiovascular
- 510(k) numbers
- K080592
- Affected lot / code info
- Batch Numbers: (1) 3000051652 3000054808 3000058733 3000062473 3000071201 3000073170 3000075123 (2) 3000076148 3000081600
Why it was recalled
Custom Tubing Sets for Extracorporeal Circulation (ECC)-During sterilization, the closed stopcock/non-vented cap configuration may prevent the flow of humidity/ethylene oxide gas into the fluid-path between the closed port and the non-vented cap therefore the Sterility Assurance Level (SAL) of the stopcock ports cannot be assured
Root cause (FDA determination)
Device Design
Action the firm took
Maquet Getinge issued recall notification letter on April 18, 2019 via FedEx 2-day delivery identifies issue, health risk and action to take: quarantine and return to Maquet/Getinge recalled product. Complete response form and return to acknowledge this recall. Questions contact Getinge Customer Service at 1 888 627 8383 (press option 2, then option 2) between the hours of 8 AM and 6 PM Eastern Standard Time. On 3/26/20 firm expanded recall to include an additional lot for one consignee,product: "BEQ TOP 25003 3/8 3/8 up to 7 lpm", item number 701063255R02, lot 3000087051.
Recalling firm
- Firm
- Maquet Cardiovascular, LLC
- Address
- 45 Barbour Pond Dr, Wayne, New Jersey 07470-2094
Distribution
- Distribution pattern
- Nationwide
Timeline
- Recall initiated
- 2019-04-18
- Posted by FDA
- 2019-05-31
- Terminated
- 2025-07-28
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #172802. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.