FDA Device Recalls

Recalls /

#173226

Product

Hudson RCI Sheridan: 1) Uncuffed 6.0 mm, Product Code 5-10412 2) Uncuffed 6.5 mm, Product Code 5-10413 3) Uncuffed 7.0 mm, Product Code 5-10414 Product Usage: Tracheal tube/airway management

FDA product code
BTRTube, Tracheal (W/Wo Connector)
Device class
Class 2
Medical specialty
Anesthesiology
510(k) numbers
K822082
Affected lot / code info
1) Uncuffed 6.0 mm, Product Code 5-10412 Lot Numbers: 73J1600363 73K1600581 73L1600347 73L1600552 73L1600850 73M1600138 73A1700264 73B1700300 73C1700688 73E1700098 73E1700712 73G1700132 73H1700327 73J1700290 73K1700065 73K1700415 73L1700169 73L1700320 73M1700250 73A1800060 73B1800348 73B1800487 73C1800349 73C1800482 73D1800464 73E1800389 73F1800171 73G1800923  2) Uncuffed 6.5 mm, Product Code 5-10413 Lot Numbers: 73M1600139 73B1700301 73E1700472 73G1700127 73K1700416 73A1800061 73B1800349 73B1800488 73D1800634 73F1800544 73G1800924  3) Uncuffed 7.0 mm, Product Code 5-10414 Lot Numbers: 73A1700100 73F1700064 73L1800406

Why it was recalled

This voluntary recall is due to reported complaints (<0.0025% of all in scope distributed product) indicating that there is an increased incidence of specific lots of the 15 mm Sheridan connector becoming disconnected from the Endotracheal tube.

Root cause (FDA determination)

Process control

Action the firm took

Teleflex sent an Urgent Medical Device Recall letter dated May 24, 2019 to customers. The letter identified the affected product, problem and actions to be taken. The letter directed customers to discontinue use, quarantine the product, and contact Teleflex for return of the product. Distributors were directed to notify their customers and request the customer return the recalled products to them for consolidation and subsequent return to Teleflex.. For questions contact your local sales representative or Customer Service at 1-866-396-2111

Recalling firm

Firm
Teleflex Medical
Address
3015 Carrington Mill Blvd, Morrisville, North Carolina 27560-5437

Distribution

Distribution pattern
Worldwide - US Nationwide Distribution

Timeline

Recall initiated
2019-05-24
Posted by FDA
2019-06-19
Terminated
2022-02-15
Status

Source: openFDA Device Recall endpoint. Recall record ID #173226. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.