Product

Hancock II T510 Bioprosthesis, Mitral, REF T510C27 - Product Usage: Replacement of malfunctioning native or prosthetic aortic and/or mitral valves.

FDA product code
DYEReplacement Heart-Valve
Device class
Class 3
Medical specialty
Cardiovascular
PMA numbers
P980043
Affected lot / code info
GTIN: 00643169027077 Serial Numbers: B654768, B694390, B818699, B887929, B896723, B896730, B914146

Why it was recalled

Fifteen affected devices were distributed worldwide, which were mislabeled with an incorrect size. In all cases, the actual surgical heart valves are larger than what was indicated on the boxes and jar labels.

Root cause (FDA determination)

Labeling mix-ups

Action the firm took

The firm initiated the recall by telephone on 05/21/2019 and followed with a letter disseminated on 06/05/2019 which requested that the consignee quarantine any unused affected product and return it to Medtronic.

Recalling firm

Firm
Medtronic Inc
Address
710 Medtronic Pkwy, Mailstop Ls245, Minneapolis, Minnesota 55432-5603

Distribution

Distribution pattern
Worldwide distribution - US Nationwide distribution in the states of TN, NY, NJ, and countries of France, Italy, Libya, Poland, Spain.

Timeline

Recall initiated
2019-05-21
Posted by FDA
2019-06-26
Terminated
2020-10-07
Status