FDA Device Recalls

Recalls /

#173456

Product

SOMATOM go.Up (Model #11061610) with syngo.CT software versions VA20A, VA20A_SP0, VA20_SP1. Product Usage: This computed tomography system is intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission data. The images delivered by the system can be used by a trained physician as an aid in diagnosis. The images delivered by the system can be used by trained staff as an aid in diagnosis, treatment preparation and radiation therapy planning. This CT system can be used for low dose lung cancer screening in high risk populations. High risk populations are as defined by professional medical societies.

FDA product code
JAKSystem, X-Ray, Tomography, Computed
Device class
Class 2
Medical specialty
Radiology
510(k) numbers
K173632
Affected lot / code info
Serial Number: 112552 112551 112555 112519 112542 112545 112541 112536 112573

Why it was recalled

The potential sporadic performance problems may result in scanning workflow interruptions and unexpected user notifications. This may cause a delay in diagnosis or patient scans.

Root cause (FDA determination)

Software design

Action the firm took

Urgent Medical Device Correction letters dated 5/24/19 were sent to customers.

Recalling firm

Firm
Siemens Medical Solutions USA, Inc
Address
40 Liberty Blvd, Malvern, Pennsylvania 19355-1418

Distribution

Distribution pattern
US Nationwide distributed in the states of AK, AL, AZ, CA, CO, CT, DC, FL, GA, IA, IL, IN, KS, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NJ, NY, OH, PA, SC, TX, VA, WA, WI, and Puerto Rico.

Timeline

Recall initiated
2019-05-28
Terminated
2023-03-31
Status

Source: openFDA Device Recall endpoint. Recall record ID #173456. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.