FDA Device Recalls

Recalls /

#173984

Product

BD Vacutainer Lithium HeparinN 75 USP Units Blood Collection Tubes, Catalog Number 367884

FDA product code
JKATubes, Vials, Systems, Serum Separators, Blood Collection
Device class
Class 2
Medical specialty
Clinical Chemistry
510(k) numbers
K901449
Affected lot / code info
UDI: (01)30382903678847

Why it was recalled

False elevation of carboxyhemoglobin (COHb) results have been reported when using venous blood gas samples collected with tubes analyzed with the IL GEM 4000 instrument.

Root cause (FDA determination)

Device Design

Action the firm took

Urgent Medical Device Correction notification letters dated 6/12/19 were sent to customers.

Recalling firm

Firm
Becton Dickinson & Company
Address
1 Becton Dr, Franklin Lakes, New Jersey 07417-1815

Distribution

Distribution pattern
Worldwide Distribution, including US Nationwide

Timeline

Recall initiated
2019-06-12
Terminated
2020-06-29
Status

Source: openFDA Device Recall endpoint. Recall record ID #173984. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.