Recalls / —
—#174208
Product
Terumo Capiox NX19 Oxygenator (East Orientation), P/N: CXOT280 Exported only to Japan for inclusion into convenience kits that are sold only within the Japanese market
- FDA product code
- DTZ — Oxygenator, Cardiopulmonary Bypass
- Device class
- Class 2
- Medical specialty
- Cardiovascular
- 510(k) numbers
- K180950
- Affected lot / code info
- Lot Numbers: XA28, XC12, XC25, XD05, XD27
Why it was recalled
Capiox NX19 Oxygenator may contain a non-functioning arterial thermistor component
Root cause (FDA determination)
Process control
Action the firm took
Terumo issued Urgent Medical Device Notification letter dated June 13 2019 via FedEx mailing executed on 6/14/2019 advising of the problem, health risk and action to take: Terumo CVS advises users to remove all affected product from inventory and return unused product to Terumo CVS following the instructions provided in this notice. Questions or concerns: Terumo CVS Customer Service: 1.800.521.2818 Monday Friday, 8 a.m. 6 p.m. ET Terumo Recall Fax: 1.410.392.7183 Terumo Recall Email: TCVSCustomerResponseElkton@terumomedical.com
Recalling firm
- Firm
- Terumo Cardiovascular Systems Corporation
- Address
- 125 Blue Ball Rd, Elkton, Maryland 21921-5315
Distribution
- Distribution pattern
- MI Foreign: Japan, Belgium
Timeline
- Recall initiated
- 2019-06-14
- Terminated
- 2020-06-11
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #174208. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.