FDA Device Recalls

Recalls /

#175139

Product

MAGNETOM Spectra magnetic resonance diagnostic devices Model # 10837643

FDA product code
LNHSystem, Nuclear Magnetic Resonance Imaging
Device class
Class 2
Medical specialty
Radiology
510(k) numbers
K121160
Affected lot / code info
Serial # 72030 72023

Why it was recalled

Potential of an open port in the quench venting system when using the optional horizontal outlet on the quench line assembly.

Root cause (FDA determination)

Device Design

Action the firm took

1. review the System Owner Manual chapter 1.3.3 Quench emergency plan and the section regarding defective vent line. 2. follow the standard emergency procedures you have in place in the event of system failures. 3.A Siemens Service Technician will conduct an inspection to verify that your system has a vertical outlet. If a vertical outlet is present, no further action is required. In the event your system is equipped with a horizontal outlet, the Siemens Service Technician will ensure that the port for the drop-off plate is adequately blanked off or sealed. 4.service organization will contact you shortly to arrange a date to perform this corrective action. Please feel free to contact our service organization for an earlier appointment at 1-800-888-7436. 5. Please review, understand, and confirm acknowledgement of the information and instructions outlined in this Customer Safety Advisory Notice. By signing this document you are acknowledging that you have read and understand the content therein. If you have any question, please contact Meredith Adams, Telephone 610-448-6461 and E-Mail: Meredith.adams@siemens-healthineers.com

Recalling firm

Firm
Siemens Medical Solutions USA, Inc
Address
40 Liberty Blvd, Malvern, Pennsylvania 19355-1418

Distribution

Distribution pattern
AL AR AZ CA CO CT DC DE FL GA GU IA ID IL IN KS KY LA MA MD ME MI MN MO MT NC NE NH NJ NY OH OK OR PA SC TN TX VA WA WI WV WY

Timeline

Recall initiated
2019-07-09
Terminated
2020-04-23
Status

Source: openFDA Device Recall endpoint. Recall record ID #175139. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.