Product

C2 Nerve Monitor, 4 Channel, 5140-508-240, UDI 07613327177329 Product Usage: The C2 Nerve Monitor System is intended for intra-operative monitoring and stimulation for localization and identification of cranial and peripheral motor and mixed motor sensory nerves during surgery.

FDA product code
ETNStimulator, Nerve
Device class
Class 2
Medical specialty
Ear, Nose, Throat
510(k) numbers
K152505
Affected lot / code info
All Serial Numbers are affected

Why it was recalled

Stryker C2 NerveMonitor's design and instructions may not optimally address usability issues related to the functionality of the device, which may result in use errors potentially causing or contributing to nerve injury.

Root cause (FDA determination)

Device Design

Action the firm took

On July 31, 2019, the firm distributed an Urgent Medical Device Recall Notification to its consignees to inform them of the recall. Customers were asked to do the following: 1. Immediately review the Recall Notification and this Business Reply Form (BRF). 2. Inform all users of the Stryker C2 NerveMonitor of the information contained in this recall notification. 3. Check all treatment areas, operating rooms and/or storage areas for affected products. 4. Determine a plan to transition to an alternative commercially available product. Document your transition date on the BRF. This transition date must be no later than January 31, 2020. 5. Complete page 2 of BRF to confirm receipt of this Notification, identify how many affected items are currently in your inventory, and provide the date you will transition to alternative product and return your Stryker C2 NerveMonitor. Upon receipt of your returned device(s), Stryker will work with your facility to issue a refund or credit. Please complete and return the BRF even if you dont have any affected product on hand. 6. Fax the completed BRF to Stryker Instruments at 8665212762, or email to kara.spath@stryker.com. 7. If you have further distributed this product, please forward copies of the Notification and this BRF to all affected locations, for each location to complete and return. Even if you have distributed all product to another location, please complete a BRF and indicate each location that received product.

Recalling firm

Firm
Stryker Instruments Div. of Stryker Corporation
Address
4100 E Milham Ave, Portage, Michigan 49002-9704

Distribution

Distribution pattern
US Nationwide distribution.

Timeline

Recall initiated
2019-07-31
Terminated
2021-03-24
Status

Source: openFDA Device Recall endpoint. Recall record ID #175420. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.