Recalls / —
—#175593
Product
BD Kiestra InoqulA Software Version 20.3.IVD automated specimen processing Catalog Number: 447202
- FDA product code
- JTC — Device, Microtiter Diluting/Dispensing
- Device class
- Class 1
- Medical specialty
- Microbiology
- Affected lot / code info
- Catalog Number: 447202 Catalog Number/UDI INO-000287 00382904472023 INO-000286 00382904472023 INO-000269 00382904472023 INO-000281 00382904472023
Why it was recalled
An anomaly is present in InoqulA / InoqulA+ software version 20.3, has the potential to cause a mismatch between a specimen and plate
Root cause (FDA determination)
Software design
Action the firm took
BD initiated a field correction via letter on July 26, 2019 and BD Technical Service has completed the correction (software upgrade) for all six (6) affected US customers. Letter identifies issue, health risk and customer action to take. BD is in the process of evaluating the log files of customers using the InoqulATM / InoqulA+TM software version 20.3 to determine if any users potentially experienced mismatch between a specimen and plate. BD will contact all customers using InoqulATM / InoqulA+TM software version 20.3 to provide the outcome of this evaluation. Contact: BD Customer/Technical Support 800-638-8663 Monday Friday between 7:00am and 7:00pm (EST) in the United States. For customers outside the US, contact your local BD representative or distributor.
Recalling firm
- Firm
- Becton Dickinson & Co.
- Address
- 7 Loveton Cir, Sparks, Maryland 21152-9212
Distribution
- Distribution pattern
- Illinois, Iowa, Missouri, New York, Ohio, Tennessee World Wide (5): Singapore, Taiwan, France, Netherlands
Timeline
- Recall initiated
- 2019-07-26
- Terminated
- 2019-12-06
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #175593. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.