Product

BD Kiestra InoqulA+ Software Version 20.3. IVD automated specimen processing Catalog Number : 447213

FDA product code
JTCDevice, Microtiter Diluting/Dispensing
Device class
Class 1
Medical specialty
Microbiology
Affected lot / code info
Catalog Number: 447213 Serial Number/UDI: INO-000290 00382904472139  INO-000263 00382904472139 INO-000264 00382904472139 20130276 00382904472139 INO-000282 00382904472139 INO-000284 00382904472139 INO-000272 00382904472139 INO-000274 00382904472139

Why it was recalled

An anomaly is present in InoqulA / InoqulA+ software version 20.3, has the potential to cause a mismatch between a specimen and plate

Root cause (FDA determination)

Software design

Action the firm took

BD initiated a field correction via letter on July 26, 2019 and BD Technical Service has completed the correction (software upgrade) for all six (6) affected US customers. Letter identifies issue, health risk and customer action to take. BD is in the process of evaluating the log files of customers using the InoqulATM / InoqulA+TM software version 20.3 to determine if any users potentially experienced mismatch between a specimen and plate. BD will contact all customers using InoqulATM / InoqulA+TM software version 20.3 to provide the outcome of this evaluation. Contact: BD Customer/Technical Support 800-638-8663 Monday Friday between 7:00am and 7:00pm (EST) in the United States. For customers outside the US, contact your local BD representative or distributor.

Recalling firm

Firm
Becton Dickinson & Co.
Address
7 Loveton Cir, Sparks, Maryland 21152-9212

Distribution

Distribution pattern
Illinois, Iowa, Missouri, New York, Ohio, Tennessee World Wide (5): Singapore, Taiwan, France, Netherlands

Timeline

Recall initiated
2019-07-26