FDA Device Recalls

Recalls /

#175829

Product

XP-CR Tibial Tray - Interlok 79 mm Item # 195276

FDA product code
JWHProsthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer
Device class
Class 2
Medical specialty
Orthopedic
510(k) numbers
K132873
Affected lot / code info
Lot Number 663820 503480 262450 308140 308150 473010 513130 320340 320590 579060 571060 938200 322150 066240 283270 322200 851360 996030 996030R 090250 841560 206870 322180 258110 279270 553900 711510 161980 161990 274450 043940 041260 663840 041170 993800 883570 883600 883590 883580 888420 374120 599170 576060 279060 712600 576050 672660 821370 600820 717210 943420 221180 141500 199770 450380 220060 502880 513120

Why it was recalled

The locking bar not fully engaging

Root cause (FDA determination)

Manufacturing material removal

Action the firm took

Risk Manager Responsibilities: 1.Review this notification and ensure that affected personnel are aware of the contents. 2.If you have affected product at your facility, assist your Zimmer Biomet sales representative and quarantine all affected product. Your Zimmer Biomet sales representative will remove the affected product from your facility. 3.Complete Attachment 1 Certificate of Acknowledgement and send to CorporateQuality.PostMarket@zimmerbiomet.com. This form will be returned even if you do not have affected products at your facility. 4.Retain a copy of the acknowledgement form with your recall records in the event of a compliance audit of your facilitys documentation. 5.If you have further questions or concerns after reviewing this notice, please call customer service at 574-371-3071 between 8:00 am and 5:00pm EST, Monday through Friday. Calls received outside of call center operating hours will receive a voicemail prompt or be transferred to an on-call representative in the event of an emergency. Alternatively, your questions may be emailed to CorporateQuality.PostMarket@zimmerbiomet.com. Surgeon Responsibilities: 1.Review this notification for awareness of the contents. 2.There are no specific patient monitoring instructions related to this recall that are recommended beyond your existing follow-up schedule. 3.A visual and audible confirmation should be made to ensure complete locking bar insertion. 4.Complete Attachment 1 Certificate of Acknowledgement and send to CorporateQuality.PostMarket@zimmerbiomet.com. 5.Retain a copy of the acknowledgement form with your recall records in the event of a compliance audit of your facilitys documentation. 6.If you have further questions or concerns after reviewing this notice, please call customer service at 574-371-3071 between 8:00 am and 5:00pm EST, Monday through Friday. Calls received outside of call center operating hours will receive a voicemail prompt or be transferred to an on-

Recalling firm

Firm
Zimmer Biomet, Inc.
Address
56 E Bell Dr, Warsaw, Indiana 46582-6989

Distribution

Distribution pattern
State NY IN OH MI GA SC FL MO WI MN ND SD NE NJ PA MD IL IA CA VA NC DC WV AR AL MS AL MI WA OR ID LA FL KS TN MA KY OK TX TX CT AZ NV NM MT WY CO UT DC WY Country: KOREA CHINA CANADA AUSTRALIA JAPAN EMEA SWEDEN ITALY

Timeline

Recall initiated
2019-08-15
Terminated
2021-04-16
Status

Source: openFDA Device Recall endpoint. Recall record ID #175829. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.