FDA Device Recalls

Recalls /

#175830

Product

XP-CR Tibial Tray - Interlok 83 mm Item # 195277

FDA product code
JWHProsthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer
Device class
Class 2
Medical specialty
Orthopedic
510(k) numbers
K132873
Affected lot / code info
Lot Number 141540 513160 571080 262460 938230 938260 966800 996040 066250 283280 663860 851390 663860R 090260 279290 376610 851380 663890 206880 374840 429310 723100 434300 469560 469560R 162000 258130 043950 320350 320330 076100 279070 600830 672690 599180 374130 576070 952400 450390 712610 717220 943430 141520 279280 376570 376590 513140

Why it was recalled

The locking bar not fully engaging

Root cause (FDA determination)

Manufacturing material removal

Action the firm took

Risk Manager Responsibilities: 1.Review this notification and ensure that affected personnel are aware of the contents. 2.If you have affected product at your facility, assist your Zimmer Biomet sales representative and quarantine all affected product. Your Zimmer Biomet sales representative will remove the affected product from your facility. 3.Complete Attachment 1 Certificate of Acknowledgement and send to CorporateQuality.PostMarket@zimmerbiomet.com. This form will be returned even if you do not have affected products at your facility. 4.Retain a copy of the acknowledgement form with your recall records in the event of a compliance audit of your facilitys documentation. 5.If you have further questions or concerns after reviewing this notice, please call customer service at 574-371-3071 between 8:00 am and 5:00pm EST, Monday through Friday. Calls received outside of call center operating hours will receive a voicemail prompt or be transferred to an on-call representative in the event of an emergency. Alternatively, your questions may be emailed to CorporateQuality.PostMarket@zimmerbiomet.com. Surgeon Responsibilities: 1.Review this notification for awareness of the contents. 2.There are no specific patient monitoring instructions related to this recall that are recommended beyond your existing follow-up schedule. 3.A visual and audible confirmation should be made to ensure complete locking bar insertion. 4.Complete Attachment 1 Certificate of Acknowledgement and send to CorporateQuality.PostMarket@zimmerbiomet.com. 5.Retain a copy of the acknowledgement form with your recall records in the event of a compliance audit of your facilitys documentation. 6.If you have further questions or concerns after reviewing this notice, please call customer service at 574-371-3071 between 8:00 am and 5:00pm EST, Monday through Friday. Calls received outside of call center operating hours will receive a voicemail prompt or be transferred to an on-

Recalling firm

Firm
Zimmer Biomet, Inc.
Address
56 E Bell Dr, Warsaw, Indiana 46582-6989

Distribution

Distribution pattern
State NY IN OH MI GA SC FL MO WI MN ND SD NE NJ PA MD IL IA CA VA NC DC WV AR AL MS AL MI WA OR ID LA FL KS TN MA KY OK TX TX CT AZ NV NM MT WY CO UT DC WY Country: KOREA CHINA CANADA AUSTRALIA JAPAN EMEA SWEDEN ITALY

Timeline

Recall initiated
2019-08-15
Terminated
2021-04-16
Status

Source: openFDA Device Recall endpoint. Recall record ID #175830. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.