Recalls / —
—#176053
Product
TRIATHLON FEMORAL DISTAL AUGMENT 10MM - SIZE 2 LEFT Catalog # 5541-A-201 - Product Usage: Triathlon Femoral Distal Augments are used with Triathlon TS Femoral Components during revision Total Knee Arthroplasty (TKA) and with Triathlon Primary PS Femoral Components during TKA to address distal femoral defects. (See attached IFU (IFU_Triathlon Total Knee_QIN 4376 Rev AB).
- FDA product code
- MBH — Prosthesis, Knee, Patello/Femorotibial, Semi-Constrained, Uncemented, Porous, Coated, Polymer/Metal/Polymer
- Device class
- Class 2
- Medical specialty
- Orthopedic
- 510(k) numbers
- K141056, K190991
- Affected lot / code info
- All lots with an expiration date on or before 31 June 2024. Lot # A3A4G A3O4G A7E4A A7S7Y AABP AABR AAS7L ABI3L ABY7A AE44H AEB7H AER7Y AGS7H AHS4D AIA7Y AII4Z ALR4L AOU3R ARG9T ASZ3O ATG4X ATU7D AVG7Z AVO9H AYG9Y AYX3Y BAE3E BAU3L BE93T BEB4L BGU4G BHL9O BIE7H BL74I BLL9A BRE9E BRO7U BRY4Y BS97E BSH7H BSO3G BTE4D DNVL DRSG DVXW DZKF FUXX FVAH GARO GAWP GBTP GJZD GKAJ GKDW GLAN GLBU GLHI GLHY HHRP HHVI HXGO HXML HXUV HYHY HZBN HZKZ HZNH IBSA IBZA IDZV IKGT IKKM ILHJ INIM INLE ITMY IXNB IYRK JBOV JDJY JEUD JIOX JIYT JJZE JKFX JKYF JLIF JOJF JWEF JWVL JYWU JZYF KEXO KMFK KMGH KMIS KOHJ KPNX KRPW KSNI KUZR KVYS MDZU MLWL MNKY MOHT MPNI MROZ S3HR SSVR TGOB U3BV ULIO USUR UZYY V4ZD VADL VGLG VHWH VJHI VJYL VPSZ VRDA VTYL VWRZ VZVT WDVE WIRA WISA WOTL WOUL WRUR WRUS WVZO WWEP WXTS WXTT WXTU WZP3 XEMZ XENA XHZU XKWY XKWZ XLRY XLRZ XSFN XSFP XSZE XSZF XVJA XVKZ XWVJ XWWZ XYGU XYGV YBJJ YBJK YCZR YCZS YFXP YFXR YGDI YGDZ YHDM YHDN
Why it was recalled
Due to device overhang issue that protrudes beyond the medial periphery of the mating femoral component
Root cause (FDA determination)
Device Design
Action the firm took
1. Inform users of this Urgent Medical Device Recall and forward this notice to all those individuals who need to be aware within your organization. 2. Immediately check all stock areas and/or operating room storage to determine if any devices from the affected product list are at your facility. 3. Quarantine and discontinue use of the recalled devices. 4. Complete and sign the enclosed Urgent Medical Device Recall Business Reply Form and fax a copy to 1-877-551-7148 or email to strykerortho4636@stericycle.com. 5. Return all affected instruments available at your location to the following address. Stryker Orthopaedics/PFA Product Returns Attn: Distribution Inventory Team 325 Corporate Drive Dock M-East Mahwah, NJ 07430 Ref. PFA 2144473
Recalling firm
- Firm
- Howmedica Osteonics Corp.
- Address
- 325 Corporate Dr, Mahwah, New Jersey 07430-2006
Distribution
- Distribution pattern
- Worldwide distribution - US Nationwide distribution including the states of AL,AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY, and countries of Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Columbia, Denmark, Finland, France, Germany, Honk Kong, India, Ireland, Iran, Italy, Israel, Japan, Kazakhstan, Korea, Kuwait, Malaysia, Mexico, Netherlands, New Zealand, Norway, Poland, Portugal, Puerto Rico, Romania, Russia, Saudi Arabia, Singapore, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, and UK.
Timeline
- Recall initiated
- 2019-08-19
- Terminated
- 2020-05-06
- Status
- —