Recalls / —
—#176817
Product
Coorad Morrey Total Elbow, Item Nos. 32810502501 32810502701 32810502900 32810503000 32810600012 32810600013 32855510796 32855510797 32855510799 32855510800 32855510801 32855510805 32855510822 32855510843 32855510844
- FDA product code
- JDC — Prosthesis, Elbow, Constrained, Cemented
- Device class
- Class 2
- Medical specialty
- Orthopedic
- 510(k) numbers
- K001989
- Affected lot / code info
- All product manufactured prior to January 2014.
Why it was recalled
Non-highly polished implants may adhere to the LDPE bag (previous packaging configuration).
Root cause (FDA determination)
Packaging change control
Action the firm took
The firm distributed Urgent Medical Device Correction letters to customers on October 10, 2019. The letter informed customers that the previous version of low density polyethylene (LDPE) bag could potentially adhere to the surface of metal devices. Please note, this is separate from the previous recall for highly polished implants packaged in the previous version of the LDPE bags. Customers were asked to take the following actions: 1. Review this notification and ensure that pertinent personnel are aware of the contents. 2. Ensure that backup devices for the affected products are available, as per standard orthopedic procedure planning, to reduce the likelihood of an extension of surgery time. 3. Examine devices thoroughly for adherent LDPE bag residue prior to use. Use a backup device if LDPE bag residue is detected. 4. Return any devices which display adherent LDPE bag residue to Zimmer Biomet, consistent with the standard instructions for use regarding the disposition of damaged components. 5. Complete the Certificate of Acknowledgement and return to the firm. 6. Retain a copy of the acknowledgement form with your field action records in the event of a compliance audit of your facility's documentation. If you have further questions or concerns regarding this recall, please call the firm's customer service at 574-371-3071 between 8:00 am and 5:00pm EST, Monday through Friday. Calls received outside of call center operating hours will receive a voicemail prompt or be transferred to an on-call representative in the event of an emergency. Alternatively, your questions may be emailed to CorporateQuality.PostMarket@zimmerbiomet.com.
Recalling firm
- Firm
- Zimmer Biomet, Inc.
- Address
- 1800 W Center St, Warsaw, Indiana 46580-2304
Distribution
- Distribution pattern
- Distributed nationwide.
Timeline
- Recall initiated
- 2019-10-10
- Terminated
- 2020-05-19
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #176817. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.