FDA Device Recalls

Recalls /

#176821

Product

NCB Proximal Lateral Tibia, "NCB-PT LAX PROX 3H TIBIA", Various Sizes, Item Nos. 0202261205 0202261207 0202261209 0202261213 0202261305 0202261307 0202261309 0202261313

FDA product code
HRSPlate, Fixation, Bone
Device class
Class 2
Medical specialty
Orthopedic
510(k) numbers
K061211
Affected lot / code info
All product manufactured prior to January 2014.

Why it was recalled

Non-highly polished implants may adhere to the LDPE bag (previous packaging configuration).

Root cause (FDA determination)

Packaging change control

Action the firm took

The firm distributed Urgent Medical Device Correction letters to customers on October 10, 2019. The letter informed customers that the previous version of low density polyethylene (LDPE) bag could potentially adhere to the surface of metal devices. Please note, this is separate from the previous recall for highly polished implants packaged in the previous version of the LDPE bags. Customers were asked to take the following actions: 1. Review this notification and ensure that pertinent personnel are aware of the contents. 2. Ensure that backup devices for the affected products are available, as per standard orthopedic procedure planning, to reduce the likelihood of an extension of surgery time. 3. Examine devices thoroughly for adherent LDPE bag residue prior to use. Use a backup device if LDPE bag residue is detected. 4. Return any devices which display adherent LDPE bag residue to Zimmer Biomet, consistent with the standard instructions for use regarding the disposition of damaged components. 5. Complete the Certificate of Acknowledgement and return to the firm. 6. Retain a copy of the acknowledgement form with your field action records in the event of a compliance audit of your facility's documentation. If you have further questions or concerns regarding this recall, please call the firm's customer service at 574-371-3071 between 8:00 am and 5:00pm EST, Monday through Friday. Calls received outside of call center operating hours will receive a voicemail prompt or be transferred to an on-call representative in the event of an emergency. Alternatively, your questions may be emailed to CorporateQuality.PostMarket@zimmerbiomet.com.

Recalling firm

Firm
Zimmer Biomet, Inc.
Address
1800 W Center St, Warsaw, Indiana 46580-2304

Distribution

Distribution pattern
Distributed nationwide.

Timeline

Recall initiated
2019-10-10
Terminated
2020-05-19
Status

Source: openFDA Device Recall endpoint. Recall record ID #176821. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.