Recalls / —
—#177043
Product
Airway Gas Option N-CAiO
- FDA product code
- CCK — Analyzer, Gas, Carbon-Dioxide, Gaseous-Phase
- Device class
- Class 2
- Medical specialty
- Anesthesiology
- 510(k) numbers
- K133576
- Affected lot / code info
- Mfg. Lot or Serial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
Why it was recalled
Displayed End Tidal Oxygen (EtO2) and Fraction of Inspired Oxygen (FiO2) values may be incorrect which could cause up to 50% measurement error in the EtO2/FiO2 values.
Root cause (FDA determination)
Process control
Action the firm took
The firm issued the Urgent Medical Device Correction Letter dated 09/27/2019. The letter describes the failure and requests replacement of the affected module with another module. If a replacement is not available, the consignee is directed to follow the instructions below to ensure that your respiratory module O2 reading is within specifications (a) before any new patient case is started and (b) for prolonged patient cases, at a minimum of once a day: 1) Connect the module to the host device; 2) Ensure the module is fully warmed up by having it powered on for minimum 20 minutes; 3) Configure the O2 reading and waveform to the host device screen as defined in the host device instructions; 4) Feed 100% O2 gas flow from the ventilator to the module and observe the O2 reading and waveform; and 5) Ensure the O2 reading is within specifications 100% +/-3% and there is no erratic behavior in the waveform. 6) If you are unable to complete the above steps successfully, remove the respiratory module from clinical use and quarantine. Contact GE Healthcare Service or your local Service Representative. If you have any questions or concerns regarding this notification, please contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative.
Recalling firm
- Firm
- GE Healthcare, LLC
- Address
- 3000 N Grandview Blvd, Waukesha, Wisconsin 53188-1615
Distribution
- Distribution pattern
- US OUS: Australia, Austria, Belgium, Bolivia, Brazil, Canada, Chile, China, Colombia, Costa Rica, Czech Republic, Finland, France, Germany, Hong Kong, India, Indonesia, Ireland, Italy, Japan, Korea (Republic of), Malaysia, Mexico, Myanmar, Netherlands, New Zealand, Peru, Poland, Republic of Serbia, Romania, Russia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, Uganda, and Unites Kingdom
Timeline
- Recall initiated
- 2019-09-27
- Posted by FDA
- 2020-01-10
- Terminated
- 2023-09-26
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #177043. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.