FDA Device Recalls

Recalls /

#178650

Product

iChemVELOCITY, Urine Chemistry System, Catalog numbers: 700-7176-001, 700-7177-001, B75502 - Product Usage: used to automate the urine chemistry analysis profile using iChem VELOCITY Urine Chemistry Strips.

FDA product code
KQOAutomated Urinalysis System
Device class
Class 1
Medical specialty
Clinical Chemistry
510(k) numbers
K101852
Affected lot / code info
UDI: 10837461009375, 10837461001751; All Part Numbers

Why it was recalled

Customers have reported incidents where two sets of results report the same Specimen Identifier (Specimen ID) with different results, different Medical Record Number (MRN) and different patient identification (demographics).

Root cause (FDA determination)

Software in the Use Environment

Action the firm took

An Urgent Medical Device Recall letter was sent during the week of 10/28/2019 to the affected customers via mail and e-mail. The notice stated the misidentification issue as follows: Customers have reported incidents where two sets of results report the same Specimen Identifier (Specimen ID) with different results, different Medical Record Number (MRN) and different patient identification (demographics)." The following actions were provided to reduce the potential for this issue until a software update is available: - Follow good laboratory practice (GLP) and ensure that a unique identification system is in place. - Ensure that specimens are labeled with a unique Specimen ID and manual orders utilize a unique Specimen ID. - Do not reuse barcode number that identify different specimens. If barcode numbers must be reused, ensure results have been released from the LIS and the instrument Work List has been cleared. - Clear all manual Orders immediately after results are released, even if barcodes are used for subsequent runs (see iQ200 or iChemVELOCITY IFU, Chaper 7, Manual Orders).

Recalling firm

Firm
Beckman Coulter Inc.
Address
250 S Kraemer Blvd, Brea, California 92821-6232

Distribution

Distribution pattern
Worldwide distributions.

Timeline

Recall initiated
2019-10-28
Terminated
2026-04-01
Status

Source: openFDA Device Recall endpoint. Recall record ID #178650. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.

— recall: iChemVELOCITY, Urine Chemistry System, Catalog numbers: 700-7176-001, 700-7177-001, B75502 - Produc · FDA Device Recalls