FDA Device Recalls

Recalls /

#179659

Product

CareFusion Alaris PCA Module, Model 8120

FDA product code
FRNPump, Infusion
Device class
Class 2
Medical specialty
General Hospital
510(k) numbers
K012383, K051641
Affected lot / code info
Software versions 12.1.0, and 9.33 and prior

Why it was recalled

1)Error 255-XX-XXX results in inability to edit settings 2)Delay options programming may result in no KVO rate/therapy interruption 3)Software errors results in no low battery alarm/infusion stopping 4)Medium priority KVO/End of Infusion alarms may result in unrecognized infusion completion 5)Custom concentration data entry errors results in concentrations being lower/higher than medication orders

Root cause (FDA determination)

Device Design

Action the firm took

On 02/04/20, Urgent Medical Device Recall Notifications were mailed and posted on the firm's website: www.bd.com/alaris-system-software-recall Two notifications were posted: 1) Software versions 9.33 and prior, and 2) software version 12.1.0. On 02/05/20. Recall letters can be found at https://www.bd.com/en-us/recall-notifications/detail-mms-20-1953 Firm will contact all customers to initiate the scheduling process for the software update. Before the update customers were asked to so the following: 1) Software Errors related to System Error 255-XX-XXX: If the error occurs while administering a critical medication(s), continue the infusion while you expedite a replacement pump if one is readily available, or restart and reprogram the PC unit. 2) KVO Rate Not Available When Using Delay Options Programming: For software versions 9.19 and prior: Set a Callback alert of "After" or "Before and After." For all software versions: Do not use Delay Options when a KVO rate is required. 3) Low Battery Alarm Failure: Whenever possible, keep the PC unit plugged into AC power. 4) Keep Vein Open (KVO) / End of Infusion Alarms Priority: Check and increase the audio volume. 5) Use Errors related to Custom Concentration Programming: Only use custom concentration when the medication label does not match any of the drug concentration selections on the programming screen. Customers can call the following phone numbers for assistance: Customer Advocacy at 888-812-3266, Clinical and Pharmacy Support Center at 858-617-1316, Recall Support Center 888-562-6018, and Technical Support 888-812-3229. On 07/29/21, Follow up Medical Device Recall Notifications were mailed to customers informing them that a new software version, v12.1.2., is available to remediate this recalls issues. Complete the acknowledgement form and contact the Recall Support Center to schedule remediation.

Recalling firm

Firm
CareFusion 303, Inc.
Address
10020 Pacific Mesa Blvd, San Diego, California 92121-4386

Distribution

Distribution pattern
Worldwide Distribution: (U.S. Nationwide) to states of: CO, SC, NY, CA, LA, TX, AK, ID, VA, DE, FL, PA, KY, KS, TN, WA, MS, IL, GA, AR, NC, NV, NE, OR, RI, WI, HI, GU, PR, MA, IA, CT, MI, AZ, ME, MT, IN, OH, WV, WY, MN, AL, NH, OK, DC, NJ, MO, SD, UT, NM, MD, VT, MP, ND, AS; and countries of: Canada, Australia, New Zealand, the Middle East, South Africa, Taiwan, Southeast Asia and the European Union.

Timeline

Recall initiated
2020-02-04
Posted by FDA
2020-03-05
Status

Source: openFDA Device Recall endpoint. Recall record ID #179659. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.