FDA Device Recalls

Recalls /

#180746

Product

Cortoss Bone Augmentation Material, Catalog Nos. 2101-0005 and 2101-0010

FDA product code
NDNCement, Bone, Vertebroplasty
Device class
Class 2
Medical specialty
Orthopedic
510(k) numbers
K080108
Affected lot / code info
Specific shipments of 2101-0005: Lots A1907006, A1908003; 2101-0010: Lots A1903054, A1906006. UDI 00808232000962, 00808232000979

Why it was recalled

Shipments were transported outside of the required refrigerated temperature conditions. Potential hazards if used include nerve damage resulting in paralysis and increased potential for embolism.

Root cause (FDA determination)

Storage

Action the firm took

On April 1, 2020, the firm distributed Urgent Medical Device Recall letters to all affected customers. Customers were informed that the product had been transported outside of the required refrigerated temperature conditions, which protect the active material components from degradation, and help maintain optimal product set time as well as optimal mechanical properties. Customers were instructed to do the following: NOTE: due to the potential hazards/risks associated with future use of product associated with these shipments, please urgently quarantine the in-scope product. 1. Immediately review the Product Recall Notification. Please forward this information to all appropriate personnel within your organization. 2. Immediately check your internal inventory to locate the product listed on the attached business reply form and list the quantity of units on the form. Each form is hospital specific and will include shipment(s) details, including customer P.O. and delivery date. 3. Use the Business Reply Form to reconcile any affected product. Complete the Business Reply Form even if there is no affected product identified. 4. Return the enclosed business reply form by email to Spine-RegulatoryActions@Stryker.com or fax to 201.575.4675 to confirm receipt of this notification. 5. Upon receipt of the completed Business Reply Form, Stryker will contact you to arrange for the return and replacement of your product(s). 6. After completion of the actions as described above, there are no additional actions that users can and should take to ensure that the product operates safely. - If you desire additional training associated with these instructions, please contact our Regulatory Compliance team at 201.749.8090 or by email at Spine-RegulatoryActions@Stryker.com. 7. Maintain awareness of this communication internally until all required actions have been completed within your facility. 8. Inform Stryker if any of the subject devices have been distributed to othe

Recalling firm

Firm
Howmedica Osteonics Corp.
Address
2 Pearl Ct, Allendale, New Jersey 07401-1611

Distribution

Distribution pattern
US distribution to California Colorado Florida Georgia Illinois Indiana Kansas Michigan Ohio Pennsylvania Texas Utah Virginia Washington Wisconsin

Timeline

Recall initiated
2020-04-01
Terminated
2020-08-27
Status