Recalls / —
—#181143
Product
Extremities Item Number: 1)113954 Hybrid Glenoid Glenoid Base, 4 MM 2) XL-115366 Comprehensive Reverse Shoulder Humeral Bearing, ArComXL, 44 MM X 41 MM, Standard 3) XL-115364 Comprehensive Reverse Shoulder Humeral Bearing, ArComXL, 44 MM X 36 MM, +3MM Standard Product Usage: is indicated for use in patients whose shoulder joint has a grossly deficient rotator cuff with severe arthropathy and/or previously failed shoulder joint replacement with a grossly deficient rotator cuff.
- FDA product code
- PHX — Shoulder Prosthesis, Reverse Configuration
- Device class
- Class 2
- Medical specialty
- Orthopedic
- 510(k) numbers
- K080642
- Affected lot / code info
- Lot Numbers: 1)597000 UDI (01)00880304462489(17)211229(10)597000 2) 502510 UDI (01)00880304475458(17)211229(10)502510 3) 744880 UDI (01)00880304475434(17)220117(10)744880
Why it was recalled
Potential presence of elevated endotoxin levels that exceed the specification limit
Root cause (FDA determination)
Process control
Action the firm took
Zimmer Biomet issued Urgent Medical Device recall letter dated April 17, 2002 to distributors with direct sales and/or field inventory via email and traceable letter sent via FedEx. Hospitals with direct sales or current inventory, as well as hospitals that have consumed products will be notified via FedEx. Surgeons that have consumed products will be notified via FedEx. Action to take: 1. Review this notification and ensure that affected team members are aware of the contents. 2. Immediately locate and quarantine affected product in your inventory. 3. Immediately return all affected product from your distributorship and from affected hospitals within your territory. a. Complete Attachment 1 Inventory Return Certification Form and send to CorporateQuality.PostMarket@zimmerbiomet.com within three (3) days. This form must be returned even if you do not have affected products available to return in your territory. b. For international returns, request an IRA by emailing zimmerbiometintlirarequests@zimmerbiomet.com c. For each return, send a copy of Attachment 1 to CorporateQuality.PostMarket@zimmerbiomet.com. CF04107 Rev. 4, Eff. Date: 16 Jul 2019 Ref. CP04102 Field Action Activities d. Include a hardcopy of Attachment 1 in each carton of your return shipment for immediate processing. e. Mark RECALL on the outside of the returned cartons. 4. Return the Additional Accounts form to CorporateQuality.PostMarket@zimmerbiomet.com. a. Review the list of hospitals and/or surgeons included with the email notification sent to your facility, which includes a list of hospitals and/or surgeons that have already been notified of this recall. b. Identify whether there are any additional hospitals and/or surgeons that Zimmer Biomet has not notified and list these accounts on the Additional Accounts form. Please provide the form in Excel format. c. If there are no additional accounts or surgeons to notify, please indicate that there are no additional accounts, or i
Recalling firm
- Firm
- Biomet, Inc.
- Address
- 56 E Bell Dr, Warsaw, Indiana 46582-6989
Distribution
- Distribution pattern
- Worldwide distribution - US Nationwide distribution and the countries of Canada, AHMADABAD, GUJARAT, BANGKOK, BELROSE NSW, BUENOS AIRES, CHILE, COSTA RICA, HIRATSUKA CITY, HUECHURABA SANTIAGO, INDIA, JAPAN, MEXICO DF, NETHERLANDS, PICHINCHA, EC, SAN JOSE, SINGAPORE, TAIWAN, THAILAND, YONGSAN-GU SEOUL.
Timeline
- Recall initiated
- 2020-04-17
- Terminated
- 2021-06-25
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #181143. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.