Recalls / —
—#181230
Product
Siemens Cios Spin VA30-mobile X-Ray system, Material # 10308194 - Product Usage: The Cios Alpha and Cios Spin are a mobile X-Ray system designed to provide X-ray imaging of the anatomical structures of patient during clinical applications.
- FDA product code
- OWB — Interventional Fluoroscopic X-Ray System
- Device class
- Class 2
- Medical specialty
- Radiology
- 510(k) numbers
- K181550
- Affected lot / code info
- Serial Numbers: 50124 50135 50131 50171 50129 50032 50033 50048 50050 50051 50073 50075 50081 50088 50098 50107 50109 50111 50120 50196 50102 50180 50136 50185 50096 50155 50091 50130 50127 50077 50078 50134 50122 50101 50092 50094 50166 50049 50192 50161 50125 50128 50132 50079 50160 50030
Why it was recalled
Main cable can be plugged or unplugged from the X10 connector which is part of the main unit (c-arm) at any time during clinical session. If the system displays the error message Err 16305 / 80, the unplugging of the main cable can cause an unsafe electrical voltage (125 V DC) at the X10 connector and result in an electrical shock if the user or other persons touch the contact pins of the X10 connector. These contact pins are accessible when the main cable is unplugged from the X10 connector
Root cause (FDA determination)
Process design
Action the firm took
Siemens Healthcare issued A Customer Safety Advisory Notice dated 4/20/20 via AX030/20/S. Letter states reason for recall, health risk and action to take: We strongly recommend that you do not disconnect the main unit from the monitor trolley when the system displays the system error message Err 16305 / 80: Confirm this error and repeat your last action until the problem has been corrected by a service engineer. Contact our service organization at 1-800-888-7436. Siemens is currently developing a solution to eliminate the root cause of this problem.
Recalling firm
- Firm
- Siemens Medical Solutions USA, Inc
- Address
- 40 Liberty Blvd, Malvern, Pennsylvania 19355
Distribution
- Distribution pattern
- US Nationwide distributions.
Timeline
- Recall initiated
- 2020-04-20
- Posted by FDA
- 2020-05-04
- Terminated
- 2021-05-10
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #181230. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.