FDA Device Recalls

Recalls /

#181432

Product

smith&nephew R3 ZERO HOLE HA COATED HEMISPHERICAL STIKTITE COATED SHELL, manufactured in the following sizes: Size REF/product number 42MM 71332242 46MM 71332246 48MM 71332248 50MM 71332250 52MM 71332252 54MM 71332254 56MM 71332256 58MM 71332258 60MM 71332260 62MM 71332262 64MM 71332264 Orthopedic implant component.

FDA product code
MBLProsthesis, Hip, Semi-Constrained, Uncemented, Metal/Polymer, Porous
Device class
Class 2
Medical specialty
Orthopedic
510(k) numbers
K070756
Affected lot / code info
71332242, Lot Numbers: 18HM20693, 18HM20693A  71332246, Lot Numbers: 19EM17729, 19EM17730, 19EM17730A  71332248, Lot Numbers: 18MM01856, 19BM08386, 19DM17884, 19DM17885, 19DM17890, 19EM11946, 19EM11947, 19DM17881, 18GM04752, 18GM04753, 18HM00807  71332250, Lot Numbers: 18JM05436, 18JM05435, 18JM05436, 19BM17964, 19DM06430, 19DM06500, 19EM03960, 18FM17594, 18GM01199, 18EM13916, 18EM13934, 18JM12796, 18JM12800, 19BM17961, 19CM15771, 19EM03956, 19EM03970, 18EM13916A, 18EM13916B, 18FM00218  71332252, Lot Numbers: 18FM02450, 18FM11167, 18GM09577, 18HM15279, 18HM15304, 18HM23477, 18HM23481, 18JM12802, 18JM12811, 18KM03172, 18LM10404, 18LM10407, 19DM21921, 19DM21927, 19EM09563, 19EM09574, 18FM01063, 18GM01228, 18GM04749, 18GM06567, 18GM17030, 18EM06114, 18EM06116, 18FM01038, 18FM00235, 18MM02061, 18MM05779, 19DM21920, 19DM21922, 19DM21925, 19DM21920A  71332254, Lot Numbers: 18FM02451, 18GM16977, 18HM15323, 18JM08634, 18JM08635, 18JM08636, 18LM10411, 18LM10411A, 18LM10411B, 18LM10414, 18MM10067, 18MM10067A, 18MM10070, 18MM10127, 18MM10127A, 18MM10144, 19BM06673, 19BM16061, 19BM16064, 19BM16066, 19BM16067, 19BM16068, 19CM04082, 19CM04084, 19CM08204, 19CM08206, 19CM08207, 19CM08209, 19CM08211, 19CM19656, 19CM19658, 19CM19658A, 19CM19703, 19CM19765, 19CM19769, 19CM19838, 19CM19838A, 19CM19839, 19CM19843, 19DM05672, 18FM16504, 18FM16506, 18LM10415, 19CM08208, 19CM19760, 19EM09578, 19EM09588, 19EM09588A, 18KM04390  71332256, Lot Numbers: 18GM07672, 18MM07589, 18MM07599, 18MM07622, 19BM12639, 18FM11173, 18FM11177, 18GM16978, 18MM07597, 19BM12636, 19EM09583, 18MM07597A  71332258, Lot Numbers: 18EM23223, 18FM02458, 18GM09584, 18GM16981, 18GM18527, 18GM18527B, 18GM18527C, 19CM11413, 19EM03998, 19EM03999, 18EM23220, 18LM17983, 18LM17984, 19BM07201, 19BM07202, 19BM07214, 19CM11417, 19DM12931, 19DM12936, 19EM04000, 19EM04001, 19CM11417A  71332260, Lot Numbers: 18HM08736, 18HM16488, 18FM02468, 8FM02468B,  18FM21552, 18HM19141, 18HM16487, 18HM16487A   71332262, Lot Numbers: 19EM11295, 19EM11296, 18MM02014  71332264, Lot Numbers: 19EM17718

Why it was recalled

A manufacturing error resulted in out of specification R3 Acetabular Shells

Root cause (FDA determination)

Process control

Action the firm took

The firm notified its consignees by letter (delivered by Fed Ex) and email beginning on 04/22/2020. Consignees are asked to inspect inventory, locate and quarantine affected devices. Affected product will be returned to Smith & Nephew. Questions or concerns regarding this recall may be directed to: FieldActions@smith-nephew.com.

Recalling firm

Firm
Smith & Nephew, Inc.
Address
1450 E Brooks Rd, Memphis, Tennessee 38116-1804

Distribution

Distribution pattern
Worldwide

Timeline

Recall initiated
2020-04-22
Status