Recalls / —
—#181984
Product
Safe-T PLUS Thora-Para 5 Fr Catheter Drainage Tray
- FDA product code
- PXI — Thoracentesis Tray
- Device class
- Class 2
- Medical specialty
- General Hospital
- Affected lot / code info
- Catalog: TPT1005SP UDI: 50885403220867; Batch: 0001345010
Why it was recalled
The Universal Drainage Set (UDS) may contain valves that leak internally which may result in the introduction of up to 30cc of air into the patient.
Root cause (FDA determination)
Under Investigation by firm
Action the firm took
All affected consignees are being notified via FedEx mailed letter and/or email beginning 6/15/2020. BD requests a sub-recall to be conducted if product has been further distributed. Consignees are asked to complete the acknowledgement form included with the notification letter and return to BD.
Recalling firm
- Firm
- Carefusion 2200 Inc 75 N Fairway Dr Vernon Hills IL 60061-1845
Distribution
- Distribution pattern
- 370 US customers and 13 distributors. Product distributed in: AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MD, MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY Customer listed, including Federal distribution, uploaded.
Timeline
- Recall initiated
- 2020-05-28
- Terminated
- 2024-09-20
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #181984. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.