FDA Device Recalls

Recalls /

#182412

Product

Beacon EUS Access System Preloaded Access 90degree Needle sterile, single-use, endoscopic ultrasound device consisting of a sharp stylet and an 18.5-gauge, echo-enhanced, tip-shaped, nitinol cannula intended to pierce tissue and then act as an access channel to facilitate guidewire placement to locations of the gastrointestinal tract.

FDA product code
FCGBiopsy Needle
Device class
Class 2
Medical specialty
Gastroenterology, Urology
510(k) numbers
K180037
Affected lot / code info
Model # DSA-090-01 UDI: 10884521722583 Lot #: F2510370X, F2510900X, F2512340X, F2512378X, F2512492X, F2512620X, F2513452X

Why it was recalled

This voluntary recall is being conducted following three customer reports of the device cannula detaching during a procedure. In one case, the detached cannula was not retrieved during the procedure. Potential risks associated with a detached cannula include delay in treatment, biliary leak, pain and retained device fragments.

Root cause (FDA determination)

Under Investigation by firm

Action the firm took

Medtronic sent a letter notification to customers dated July 06, 2020 recalling effected lots of product. This letter instructed customers to discontinue use of the product and quarantine effected inventory. Customers were also instructed to return effected product and complete recall response forms accordingly.

Recalling firm

Firm
Covidien Llc
Address
15 Hampshire St, Mansfield, Massachusetts 02048-1113

Distribution

Distribution pattern
world wide distribution

Timeline

Recall initiated
2020-06-26
Posted by FDA
2020-08-05
Status