Recalls / —
—#183369
Product
Alere iScreen Single Drug Detection (AMT Item Number/ AMT Description): X11-IS1 BUP-DX/ "Alere iScreen" Dx Single Dip Card Single Drug Detection: Buprenorphine (BUP)", X11-IS1 BAR-DX /"Alere iScreen" Dx Single Dip Card Single Drug Detection: Barbiturate (BAR)", X11-IS1 OXY-DX / "Alere iScreen Dx Single Dip Card Single Drug Detection: Oxycodone (OXY)", X11-IS1 BZO-DX / "Alere iScreen" Dx Single Dip Card Single Drug Detection: Benzodiazepine (BZO300)", X11-IS1 MET-DX / "Alere iScreen Dx Single Dip Card Single Drug Detection: Methamphetamine (MET)", X11-IS1 AMP-DX /"Alere iScreen Dx Single Dip Card Single Drug Detection: Amphetamine (AMP)", X11-IS1 MDM-DX / "Alere iScreen Dx Single Dip Card Single Drug Detection: Ecstasy (MDMA)", X11-IS1 MTD-DX /"Alere iScreen Dx Single Dip Card Single Drug Detection: Methadone (MTD)", X11-IS1 MOP-DX/"Alere iScreen Dx Single Dip Card Single Drug Detection: Morphine (MOP)", X11-IS1 THC-DX/"Alere iScreen Dx Single Dip Card Single Drug Detection: Marijuana (THC)", X11-IS1 TCA-DX/ "Alere iScreen Dx Single Dip Card Single Drug Detection: Tricyclic Antidrepessants (TCA)", X11-IS1 PCP-DX /"Alere iScreen Dx Single Dip Card Single Drug Detection: Phencyclidine (PCP)",
- FDA product code
- DIO — Enzyme Immunoassay, Cocaine And Cocaine Metabolites
- Device class
- Class 2
- Medical specialty
- Clinical Toxicology
- 510(k) numbers
- K022335, K082508
- Affected lot / code info
- ALL Lots Within Expiry
Why it was recalled
Various immunoassay enzymes failed product performance, thereby the products do not consistently detect all calibrator and secondary compounds at the levels stated in the labeling.
Root cause (FDA determination)
Process control
Action the firm took
A Field Corrective Action Notice letter was sent to affected consignees. The Customer notifications commenced on 19-Aug-2020. The consignees are requested to return all non-expired product in their possession to the recalling firm. Method of Notification, 1st contact: Written field corrective action notification letters will be sent via certified mail with delivery tracking to each consignee along with a pre-paid return label to return any product remaining in stock. 2nd & 3rd contact: Telephone, follow up email, or written mail communication will be used as necessary.
Recalling firm
- Firm
- Ameditech Inc
- Address
- 9940 Mesa Rim Rd, San Diego, California 92121-2910
Distribution
- Distribution pattern
- US - Nationwide OUS - Australia, France, Germany, New Zealand, Singapore and Spain
Timeline
- Recall initiated
- 2020-08-18
- Terminated
- 2024-09-30
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #183369. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.