FDA Device Recalls

Recalls /

#183734

Product

System, Nuclear Magnetic Resonance Imaging - Product Usage: is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times.

FDA product code
LNHSystem, Nuclear Magnetic Resonance Imaging
Device class
Class 2
Medical specialty
Radiology
510(k) numbers
K091536, K163331
Affected lot / code info
Model: 5741000-23 and 5396469

Why it was recalled

Potential electrical arc flash within the Power Gradient & RF (PGR) Cabinet.

Root cause (FDA determination)

Process change control

Action the firm took

On 09/08/20 the firm sent a Customer Letter to its consignees with the following instructions: You can continue to use the system. In the event the PGR cabinet needs to be serviced, power down the PGR at the Main Disconnect Panel (MDP) then follow MDP Lockout/Tagout from the current released service methods and documentation. GE Healthcare will correct all affected products at no cost to you. A GE Healthcare representative will contact you to arrange for the correction

Recalling firm

Firm
GE Healthcare, LLC
Address
3000 N Grandview Blvd, Waukesha, Wisconsin 53188-1615

Distribution

Distribution pattern
Worldwide distribution - US Nationwide distribution including in the states of FL, KY, NC, NJ, NM, NY, TX, UT and WA. The countries of Canada, Hungary, India, Korea, China, Russia, Taiwan, Puerto Rico and Ubekistan.

Timeline

Recall initiated
2020-09-08
Terminated
2021-04-28
Status

Source: openFDA Device Recall endpoint. Recall record ID #183734. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.