FDA Device Recalls

Recalls /

#183795

Product

HEALICOIL RSB SA 5.5MM W/1 UT & 1 UB BL- Absorbable Suture anchor Product Number: 72203801

FDA product code
MAIFastener, Fixation, Biodegradable, Soft Tissue
Device class
Class 2
Medical specialty
Orthopedic
510(k) numbers
K151105
Affected lot / code info
Batch Numbers: 2013918 2016156 2017774 2018804 2021456 2023227 2024815 2025049 2027438 2029400 2029401 2033291 2033290 2034431 2034432 2034499 2034500 2034501 2034502 2035698 2037194 2039456 2045894 2048429 2048430 2051045 2051046 2049950 2049951 2053927 2056086 2056434

Why it was recalled

Product packaging improper or incomplete seal of the pouch surrounding the sterile product may result in a breach of the sterile barrier created by the pouch

Root cause (FDA determination)

Equipment maintenance

Action the firm took

Smith & Nephew issued Urgent Medical Device Recall letter (R-2020-19) FSN on September 9, 2020 via email or overnight mail to customers that received affected products. Letter states reason for recall, health risk, and action to take: Required Actions: 1. Please inspect your inventory and locate any devices from the listed product and batch numbers enclosed with this Urgent Medical Device Recall Notice, and quarantine them immediately. a. If you are a sales rep or distributor, you must notify your customers of the field action and ensure that the required actions are complete. b. Include the customers account number and name in the table below. 2. If you have no product to return, please put an X in the appropriate location below. 3. If you have product to return, please list the item, batches and quantities that you are returning in the appropriate boxes below. 4. Complete the remainder of the form, sign and send to FieldActions@smith-nephew.com. Please Note even if you have no product to return, this form must be completed, signed and returned. 5. Once the form is received by Smith & Nephew, you will be sent a Return Authorization (RA) number. If you have any questions or concerns regarding this recall please contact FieldActions@smith-nephew.com.

Recalling firm

Firm
Smith & Nephew, Inc.
Address
150 Minuteman Rd, Andover, Massachusetts 01810-1031

Distribution

Distribution pattern
Nationwide Foreign: AE, AT, AU, BE, BM, BR, CA, CH, CL, CN, CO, CY, CZ, DE, DK, DO, EC, EE, ES, FI, FR, GB, GE, GR, HK, HU, IE, IL, IN, IT, JP,KR, LV,MY, NL, NO, PA,PE, PL, PR, PT, RU, SE, SG, SI, TH, TR, TW, UA, & ZA

Timeline

Recall initiated
2020-09-09
Status