Recalls / —
—#183898
Product
ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 22 mm, Neutral, O.D. Cup with Spacers, 57 mm Item Number: 00-8065-554-22
- FDA product code
- JDI — Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Cemented
- Device class
- Class 2
- Medical specialty
- Orthopedic
- Affected lot / code info
- All lots
Why it was recalled
Sterilant (hydrogen peroxide) used in acetabular cup hip prosthesis component was not evaluated to confirm the biocompatibility of the residual sterilant. Cannot rule out the potential for adverse tissue reactions.
Root cause (FDA determination)
Device Design
Action the firm took
Zimmer Biomet issued URGENT MEDICAL DEVICE RECALL AND NOTICE OF US FIELD DISCONTINUATION letter to Distributors. Risk Manages, Physicianson/ 9/24/20 stating reason for recall, heath risk and action to take: 1.Review this notification and ensure that affected personnel are aware of the contents. 2. If you have affected product at your facility, assist your Zimmer Biomet sales representative and quarantine all affected product. Your Zimmer Biomet sales representative will remove the affected product from your facility. 3. Complete Attachment 1 Certificate of Acknowledgement and send to CorporateQuality.PostMarket@zimmerbiomet.com. This form will be returned even if you do not have affected products at your facility. 4. Retain a copy of the acknowledgement form with your recall records in the event of a compliance audit of your facilitys documentation. 5. If you have further questions or concerns after reviewing this notice, please call customer service at 574-371-3071 between 8:00 am and 5:00pm EST, Monday through Friday. Calls received outside of call center operating hours will receive a voicemail prompt or be transferred to an on-call representative in the event of an emergency. Alternatively, your questions may be emailed to CorporateQuality.PostMarket@zimmerbiomet.com. Surgeon Responsibilities: 1. Review this notification for awareness of the contents. 2. There are no specific patient monitoring instructions related to this recall that are recommended beyond your existing follow-up schedule.
Recalling firm
- Firm
- Zimmer Biomet, Inc.
- Address
- 1800 W Center St, Warsaw, Indiana 46580-2304
Distribution
- Distribution pattern
- US Nationwide
Timeline
- Recall initiated
- 2020-09-24
- Terminated
- 2022-05-20
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #183898. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.