Recalls / —
—#185055
Product
GE NM/CT 870 CZT system Model Number H3906CW - Product Usage: intended for general Nuclear Medicine imaging procedures for detection of radioisotope tracer uptake in the patient body.
- FDA product code
- KPS — System, Tomography, Computed, Emission
- Device class
- Class 2
- Medical specialty
- Radiology
- 510(k) numbers
- K153402
- Affected lot / code info
- Serial numbers: 87CY63028, 87CZ63044, 87CZ63045, 87CZ63042, 87CY63025, 87CY63029, 87CY63027, 87CY63030, 87CZ63046, 87CY63034, 87CY63032, 87CY63033
Why it was recalled
GE Healthcare has become aware that rotor bearing screws were found loose on one detector in the field, leading to the release of one of the four rotor bearings in the detector.
Root cause (FDA determination)
Process control
Action the firm took
The firm initiated their action by letter to the consignees on 12/10/2020. The letter explained the issue and stated that GE would inspect, and if required, correct the affected units. Questions and concerns may be directed to GE Healthcare Service at 1-800-437-1171 or the consignee's local Service Representative.
Recalling firm
- Firm
- GE Healthcare, LLC
- Address
- 3000 N Grandview Blvd, Waukesha, Wisconsin 53188-1615
Distribution
- Distribution pattern
- Worldwide
Timeline
- Recall initiated
- 2020-12-10
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #185055. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.